FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 6753303 · Received July 31, 2017

Report

Report Number
2531779-2017-16529
Event Type
Malfunction
Date Received
July 31, 2017
Report Date
July 1, 2017
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

FOLLOW-UP #1: DATE OF SUBMISSION 10/30/2017 - DEVICE EVALUATION: IN Q3 2017, THERE WERE 9 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Additional Manufacturer Narrative · 0

OF THE 46 ALLEGATIONS OF A MALFUNCTION, 33 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, 4 WERE FOUND TO BE MALFUNCTIONING DUE TO A TORN TRANSCEIVER FLEX CABLE. THIS REPORT IS PROCESSED UNDER EXEMPTION NUMBER E2015053. THIS MDR REPORT IS PART OF THE 227 PILOT PROGRAM.

Description of Event or Problem · 0

THIS REPORT SUMMARIZES 46 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 5-74 YEARS OF AGE AND BETWEEN 23 AND 240 POUNDS. OF THE REPORTED PATIENTS, 22 WERE MALE, 24 WERE FEMALE.