FDA Adverse Event Malfunction Summary report: N

BROMELIN FOR ERYTYPE, 10 ML

MDR report key: 6747023 · Received July 27, 2017

Report

Report Number
9610824-2017-00049
Event Type
Malfunction
Date Received
July 27, 2017
Date of Event
June 29, 2017
Report Date
August 11, 2017
Manufacturer
BIO-RAD MEDICAL DIAGNOSTICS GMBH
Product Code
KSZ
UDI-DI
07611969965755
PMA / PMN Number
BK040070
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS IS OUR FINAL REPORT ON THIS INCIDENT.

Additional Manufacturer Narrative · 1

THIS IS OUR INITIAL REPORT ON THIS INCIDENT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A DISCREPANCY WHEN TYPING ON TWO TANGO OPTIMO INSTRUMENTS. RH NEGATIVE SAMPLES WERE BEING READ AS POSITIVE. THE CUSTOMER SUSPECTED THE BROMELIN FOR ERYTYPE TO BE THE CAUSE FOR THE FALSE POSITIVE RESULTS. THE CUSTOMER SENT BACK THE ALLEGEDLY DEFECTIVE PRODUCT BROMELIN FOR ERYTYPE FOR INVESTIGATIONAL TESTING BUT AT THE TIME THE ALLEGEDLY DEFECTIVE PRODUCT ARRIVED AT BMD IT WAS ALREADY EXPIRED, WAS CONTAMINATED AND EVEN MOULDY. THEREFORE OUR QUALITY CONTROL LABORATOY TESTED THEIR RETAINED SAMPLE BROMELIN FOR ERYTYPE, #8701110-00. THE ERYTYPE S ABD+REV. A1B #8646110, THAT WAS USED BY THE CUSTOMER, WAS ALSO USED IN THIS TEST. 12 SAMPLES WERE TESTED AND ALL REACTIONS WERE AS EXPECTED. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVE INFLUENCES ON THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT OF BROMELIN FOR ERYTYPE. BOTH AFFECTED TANGO OPTIMO WERE INSPECTED BY OUR FIELD SERVICE ENGINEER WITH NO INDICATION FOR A MALFUNCTION. BOTH INSTRUMENTS ARE WORKING WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ABOUT UNEXPECTED POSITIVE RESULTS OF KNOWN RH D NEGATIVE SAMPLES BY ERYTYPE S ABD+REV A1,B ON HIS BOTH TANGO OPTIMO INSTRUMENTS. THE CUSTOMER WAS SUSPECTING THAT THE BROMELIN FOR ERYTYPE+ERYTYPE S (LOT # 8701110-00) COULD BE THE CAUSE FOR THE FALSE POSITIVE RESULTS AND RETURNED THE SUPPOSEDLY DEFECTIVE PRODUCT FOR INVESTIGATIONAL TESTING. BUT ON DATE OF RECEIPT ((B)(6)2017) THE COMPLAINT MATERIAL WAS ALREADY EXPIRED (EXP.: 07/04/2017). OUR QUALITY CONTROL IS PLANNING FURTHER TESTING, THE INVESTIGATIONS ARE ONGOING. BOTH AFFECTED TANGO OPTIMO INSTRUMENTS WERE INSPECTED BY ONE OF OUR FIELD SERVICE ENGINEERS WITH NO INDICATION FOR AN INSTRUMENT MALFUNCTION. THE INSTRUMENTS WERE CONFIRMED TO OPERATE WITHIN SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
526953 BROMELIN FOR ERYTYPE, 10 ML BROMELIN FOR ERYTYPE KSZ BIO-RAD MEDICAL DIAGNOSTICS GMBH 8701110-00 07611969965755

Patients

Seq Age Sex Outcome Treatment
1