AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2017-00408
- Event Type
- Malfunction
- Date Received
- July 27, 2017
- Date of Event
- June 27, 2017
- Report Date
- July 27, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
ON 25 JULY 2017 THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEMS AND COMMENTED AS FOLLOWS: THE TERMINAL IS SCREWED AND FIXED INSIDE THE CUP. WE HAVE UNSCREWED IT USING A NEW SCREWDRIVER IN COMBINATION WITH A ROD. LOOKING AT THE TERMINAL SUFACE WE NOTICE THAT THE BALINIT-C COATING IS SLIGHTLY SCRATCHED TOGETHER WITH THE THREADED CONNECTION TIP. ALSO ON THE IMPLANTS WE NOTICED BALINIT-C FRAGMENTS AND THREADED HOLE SCRATCHES. THESE SHOULD BE THE ROOT CAUSE OF THE COMPLAINTS. BATCH REVIEWS PERFORMED ON 27 JULY 2017. LOT 1653999: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13 FEBRUARY 2017. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, OTHER (B)(4) SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. VERSAFIT CUP CC TRIO ACETABULAR SHELL Ø 54, CODE 01.26.45.0054, LOT. 168174 (K103352) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28 MARCH 2017. EXPIRATION DATE: 2022-03-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
DURING SURGERY, THE TERMINAL CUP IMPACTOR FOR METAL BACK CC STUCK INTO THE CUP AND CAN'T BE DISCONNECT. HE USED ANOTHER CUP (SAME REF.) TO COMPLETE THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 526560 | AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMENT FOR HIP | LXH | MEDACTA INTERNATIONAL SA | 1653999 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Other |