FDA Adverse Event Malfunction Summary report: N

19 G X 1 1/2 IN. BD NOKOR¿ FILTER NEEDLE

MDR report key: 6746107 · Received July 27, 2017

Report

Report Number
1911916-2017-00093
Event Type
Malfunction
Date Received
July 27, 2017
Date of Event
February 7, 2017
Report Date
November 15, 2017
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
GAA
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: NO NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCH. DHR REVIEW: ASSEMBLY BATCH 4181680 HAD 77 VISUAL INSPECTIONS PERFORMED ON (B)(4) PARTS WITH NO DEFECTS NOTED. THE ID INSPECTION SYSTEM WAS CHALLENGED 38 TIMES USING (B)(4) PARTS WITH ZERO FAILURES RECORDED. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTION: DATE RECEIVED BY MANUFACTURER: 06/06/2017. INVESTIGATION: RESULTS: NO SAMPLE OR PHOTO PROVIDED NEW INFORMATION. THIRTY-NINE SHIELD REMOVAL FORCE TESTS WERE PERFORMED ON 195 PARTS WITH A MINIMUM REMOVAL FORCE OF 2.1 LBS., A MAXIMUM REMOVAL FORCE OF 5.7 LBS., AND AN AVERAGE OF 3.49 LBS. THE SPECIFICATION LIMITS ARE 0.5 ¿ 12.0 LBS. WITH AN AVERAGE < 9.0 LBS. TWENTY CANNULA REMOVAL FORCE TESTS WERE PERFORMED ON 100 PARTS WITH A MINIMUM REMOVAL FORCE OF 18 LBS., A MAXIMUM REMOVAL FORCE OF 40 LBS., AND AN AVERAGE OF 27.8 LBS. THE SPECIFICATION LIMIT FOR A 19 GAUGE NEEDLE IS A MINIMUM REMOVAL FORCE OF 16 LBS. ROOT CAUSE: BASED ON THE INVESTIGATION CONCLUSION, BD MPS WAS NOT ABLE TO CONFIRM THE SYMPTOM PERCEIVED BY THE CUSTOMER OR CORRELATE THIS SYMPTOM WITH A POTENTIAL CAUSE LINKED TO BD PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CAP AND NEEDLE POPPED OFF OF A 19 G X 1 1/2 IN. BD NOKOR¿ FILTER NEEDLE BEFORE USE. NO INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
528827 19 G X 1 1/2 IN. BD NOKOR¿ FILTER NEEDLE FILTER NEEDLE GAA BD MEDICAL (BD WEST) MEDICAL SURGICAL 4181650

Patients

Seq Age Sex Outcome Treatment
1 Other