FDA Adverse Event Malfunction Summary report: N

CSK-TC ELECTRODE

MDR report key: 6745790 · Received July 26, 2017

Report

Report Number
3006630150-2017-02656
Event Type
Malfunction
Date Received
July 26, 2017
Date of Event
June 28, 2017
Report Date
June 28, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250011524
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: CSK-TC10-3M LOT #: 041617 DESCRIPTION: CSK 100 MM TC ELECTRODE WITH 3M CABLE.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION PERFORMED ON ELECTRODE LOT# 040612 SHOWED THAT THE ELECTRODE SHAFT WAS BROKEN OFF AT THE HUB. DEVICE EVALUATION PERFORMED ON ELECTRODE LOT# 041617 SHOWED THE THERMOCOUPLE HAS MIGRATED FROM TIP TO A DIFFERENT POSITION ON THE SHAFT. THE ELECTRODE SHAFT WAS BENT, DUE TO EXTERNAL MECHANICAL FORCE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT A USED ELECTRODE WAS FOUND TO HAVE A RUSTY WIRE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT A USED ELECTRODE WAS FOUND TO HAVE A RUSTY WIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
525077 CSK-TC ELECTRODE ELECTRODES GXI COSMAN MEDICAL CSK-TC10-3M 040612 00813250011524

Patients

Seq Age Sex Outcome Treatment
1