FDA Adverse Event Malfunction Summary report: N

TRIOS ACP TABLE BASE

MDR report key: 6745241 · Received July 26, 2017

Report

Report Number
2921578-2017-00006
Event Type
Malfunction
Date Received
July 26, 2017
Date of Event
July 17, 2017
Report Date
September 22, 2017
Manufacturer
MIZUHO ORTHOPEDIC SYSTEMS, INC.
Product Code
JEA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONDUCTED A PREVENTATIVE MAINTENANCE WITH NO PROBLEM FOUND, OPERATOR ERROR. ACCOUNT MANAGER PROVIDED IN-SERVICE TO RE-EDUCATED THE DOCTOR AND STAFF ON PROPER USE OF TABLE.

Description of Event or Problem · 1

WITH PATIENT ON THE TRIOS TABLE, THE OPERATING ROOM STAFF AND SURGEON PERFORMED 180 ROTATION. WHILE LIFTING THE ROTATION LEVER, THE SURGEON BELIEVED HE SAW THE AMBER LIGHT ON THE FOOT END FLASH BEFORE RELEASING THE HANDLE ALL THE WAY DOWN. AFTER A SUCCESSFUL ROTATION, THE STAFF WENT TO PREP FOR SURGERY AND THE SURGEON RELEASED THE ROTATION TO CHECK THE PROBLEM AGAIN. WHILE DOING SO, THE PATIENTS UPPER BODY FELL TO THE FLOOR AS THE TABLE UNLOCKED AND TILTED LATERALLY. THERE WAS A SAFETY STRAP POSITIONED NEAR THE PATIENT'S BUTTOCKS. NO INJURY HAS BEEN REPORTED.

Description of Event or Problem · 1

WITH PATIENT ON THE TRIOS TABLE, THE OR STAFF AND SURGEON PERFORMED 180 ROTATION. WHILE LIFTING THE ROTATION LEVER, THE SURGEON BELIEVED HE SAW THE AMBER LIGHT ON THE FOOT END FLASH BEFORE RELEASING THE HANDLE ALL THE WAY DOWN. AFTER A SUCCESSFUL ROTATION, THE STAFF WENT TO PREP FOR SURGERY AND THE SURGEON RELEASED THE ROTATION TO CHECK THE PROBLEM AGAIN. WHILE DOING SO, THE PATIENTS UPPER BODY FELL TO THE FLOOR AS THE TABLE UNLOCKED AND TILTED LATERALLY. THERE WAS A SAFETY STRAP POSITIONED NEAR THE PATIENT'S BUTTOCKS. NO INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
524263 TRIOS ACP TABLE BASE OPERATING TABLE JEA MIZUHO ORTHOPEDIC SYSTEMS, INC. 7803AP

Patients

Seq Age Sex Outcome Treatment
1 Other