FDA Adverse Event Other Summary report: N

AIRSHIELDS

MDR report key: 674200 · Received January 30, 2006

Report

Report Number
674200
Event Type
Other
Date Received
January 30, 2006
Date of Event
October 6, 2005
Report Date
January 30, 2006
Manufacturer
DRAEGER MEDICAL, INC.
Product Code
FMZ
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

INFANT IN ISOLETTE SET AT 26 DEGREES CELSIUS. ISOLETTE WAS UNPLUGGED FOR TRANSFER OF INFANT AND UNIT RESET TO 35 DEGREES CELSIUS WHEN PLUGGED BACK IN. INFANT NOTED TO BE TACHYPNEIC AND TACHYCARDIC WITH AXILLARY TEMPERATURE OF 38.5. STAFF UNAWARE OF AUTOMATIC RESET TO 35 DEGREES WHEN THERE IS LOSS OF POWER OR UNIT IS UNPLUGGED. THE UNIT WILL STORE THE SET TEMPERATURE FOR A 15 MINUTE PERIOD BEFORE DEFAULT TO 35 DEGREES CELSIUS. THERE WAS NO HARM TO PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AIRSHIELDS INCUBATOR FMZ DRAEGER MEDICAL, INC. C-2000 *

Patients

Seq Age Sex Outcome Treatment
1 13 DA