FDA Adverse Event Injury Summary report: N

COCR BALL HEAD 12/14 Ø 32 SIZE L +3.5

MDR report key: 6740185 · Received July 25, 2017

Report

Report Number
3005180920-2017-00406
Event Type
Injury
Date Received
July 25, 2017
Date of Event
June 28, 2017
Report Date
July 25, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805172
PMA / PMN Number
K072857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS THE SECOND REVISION SURGERY UNDERWENT BY THE PATIENT. ON (B)(6) 2017 THE PATIENT HAD A REVISION OF NON-MEDACTA PRODUCTS DUE TO INFECTION. ON 21 JULY 2017 THE MEDICAL AFFAIRS PERFORMED A CLINICAL EVALUATION BASED ON THE AVAILABLE PICTURES AND COMMENTED AS FOLLOWS: HIP DISLOCATION FEW MONTHS AFTER REVISION HIP ARTHROPLASTY FOLLOWING INFECTION. A NEW HEAD, IN THE FINAL OPERATION, TWO SIZES LONGER THAN THE PREVIOUS ONE WAS USED TO RESTORE SOFT TISSUE TENSION AND POSSIBLY LEG LENGTH. THIS MAY BE THE CONSEQUENCE OF STEM SUBSIDENCE (A POSSIBLE SEQUELA AFTER INFECTION) OR FOR INSUFFICIENT TENSION ACHIEVED DURING THE FIRST REVISION SURGERY: THE FORMER IMAGES ARE NOT AVAILABLE. THERE IS NO REASON TO SUSPECT A FAULTY DEVICE. BATCH REVIEWS PERFORMED ON 25 JULY 2017. LOT 152724: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30 OCTOBER 2015. EXPIRATION DATE: 2020-10-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD AND THIS IS THE SECOND SIMILAR EVENT REPORTED ON ITEMS OF THE SAME LOT. VERSAFITCUP FLAT PE HC LINER Ø 32 / F, CODE 01.26.3248HCT, LOT. 164467 (K103352) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 05 OCTOBER 2016. EXPIRATION DATE: 2021-09-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY DUE TO HIP LUXATION. HEAD AND LINER WERE EXPLANTED. NO TRAUMA WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
522228 COCR BALL HEAD 12/14 Ø 32 SIZE L +3.5 METALLIC BALL HEAD JDI MEDACTA INTERNATIONAL SA 152724 07630030805172

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention