IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2017-00254
- Event Type
- Malfunction
- Date Received
- July 25, 2017
- Date of Event
- June 28, 2017
- Report Date
- June 28, 2017
- Manufacturer
- MEDTRONIC XOMED INC.
- Product Code
- HBE
- UDI-DI
- 00613994933713
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
CONCOMITANT MEDICAL PRODUCTS: 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC); SERIAL NUMBER - (B)(4); LOT NUMBER - 209489570; MANUFACTURED DATE - APR/24/2015; UDI NUMBER - (B)(4); 510K - K081475 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC); SERIAL NUMBER - (B)(4); LOT NUMBER - 209489570; MANUFACTURED DATE - APR/24/2015; UDI NUMBER - (B)(4); 510K - K081475. THE DRILLS HAVE BEEN RECEIVED. THE PRODUCT ANALYSES HAVE NOT YET BEEN PERFORMED. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 87 DEGREES F AT T1 AND 83 DEGREES F AT T2. THE HANDPIECE WAS SUCCESSFULLY TESTED TO MANUFACTURING SPECIFICATIONS. THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 83 DEGREES F AT T1 AND 82 DEGREES F AT T2. THE HANDPIECE WAS SUCCESSFULLY TESTED TO MANUFACTURING SPECIFICATIONS. THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 87 DEGREES F AT T1 AND 82 DEGREES F AT T2. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE CUSTOMER REPORTED THAT DURING USE, THREE INDIGO HANDPIECE/DRILL HANDLES GOT TOO HOT TO HOLD. A DAMP SWAB WAS WRAPPED AROUND THE HANDLES. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 520443 | IPC® HANDPIECE - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED INC. | 1845000 | 209455624 | 00613994933713 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |