FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - INDIGO¿ DRILL

MDR report key: 6739596 · Received July 25, 2017

Report

Report Number
1045254-2017-00254
Event Type
Malfunction
Date Received
July 25, 2017
Date of Event
June 28, 2017
Report Date
June 28, 2017
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00613994933713
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC); SERIAL NUMBER - (B)(4); LOT NUMBER - 209489570; MANUFACTURED DATE - APR/24/2015; UDI NUMBER - (B)(4); 510K - K081475 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC); SERIAL NUMBER - (B)(4); LOT NUMBER - 209489570; MANUFACTURED DATE - APR/24/2015; UDI NUMBER - (B)(4); 510K - K081475. THE DRILLS HAVE BEEN RECEIVED. THE PRODUCT ANALYSES HAVE NOT YET BEEN PERFORMED. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 87 DEGREES F AT T1 AND 83 DEGREES F AT T2. THE HANDPIECE WAS SUCCESSFULLY TESTED TO MANUFACTURING SPECIFICATIONS. THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 83 DEGREES F AT T1 AND 82 DEGREES F AT T2. THE HANDPIECE WAS SUCCESSFULLY TESTED TO MANUFACTURING SPECIFICATIONS. THE 1845000 (SERIAL (B)(4)): THE PRODUCT ANALYSIS INDICATES THE DEVICE WAS RECEIVED IN GOOD CONDITION. THE REPORTED ISSUE OF OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS ARE AS FOLLOWS: 87 DEGREES F AT T1 AND 82 DEGREES F AT T2. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING USE, THREE INDIGO HANDPIECE/DRILL HANDLES GOT TOO HOT TO HOLD. A DAMP SWAB WAS WRAPPED AROUND THE HANDLES. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
520443 IPC® HANDPIECE - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 209455624 00613994933713

Patients

Seq Age Sex Outcome Treatment
1