FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 6739587
·
Received July 25, 2017
Report
- Report Number
- 3004123209-2017-00781
- Event Type
- Malfunction
- Date Received
- July 25, 2017
- Date of Event
- July 17, 2017
- Report Date
- August 18, 2017
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K014067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 0
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).
Additional Manufacturer Narrative · 0
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). ROUTINE TESTING CONFIRMED THAT THE PAD-PAK WAS FIRST WAS INSTALLED BY THE CUSTOMER ON (B)(6) 2010. BETWEEN (B)(6) 2016 THE DEVICE RECORDS MULTIPLE MANUAL POWER UPS OF 10 MINUTE DURATION. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
Description of Event or Problem · 0
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE OBSERVED SWITCHING ON AUTOMATICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 520219 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |