HEARTSINE SAMARITAN 350P AND PAD PAK
Report
- Report Number
- 3004123209-2017-00728
- Event Type
- Malfunction
- Date Received
- July 24, 2017
- Date of Event
- July 5, 2017
- Report Date
- August 30, 2017
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- 160008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PATIENT
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE DEVICE HISTORY RECORDS FOR THE RETURNED SAM 350P DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TESTS HAD BEEN SUCCESSFULLY COMPLETED PRIOR TO THE DESPATCH OF THE SAM 350P FROM HEARTSINE TECHNOLOGIES, BELFAST ON THE 20TH OF MAY 2014. THE HISTORY LOG FOR THIS DEVICE SHOWED THAT THE PAD-PAK WAS FIRST INSTALLED ON (B)(6) 2014 AND THAT IT HAD PERFORMED ALL SELF-TESTS UP TO AND INCLUDING THE LAST LOG ENTRY ON (B)(6) 2015. THE RETURNED PAD-PAK WAS INSTALLED AND THE DEVICE FAILED TO POWER ON. THE DEVICE WAS DISASSEMBLED TO INVESTIGATE AND UPON VISUAL INSPECTION THE BATTERY WAS OBSERVED TO BE SEVERELY BLOWN WITH SILVER PARTICLES OBSERVED ACROSS THE COMPONENTS ON THE UPPER AND LOWER SIDE OF THE PCB. AFTER EXTENSIVE INVESTIGATION, A TRANSISTOR WHICH IS PART OF THE STATUS LED CIRCUITRY WAS FOUND TO HAVE FAILED WHICH LED TO THE GREEN STATUS LED REMAINING CONSTANTLY LIT. THIS TRANSISTOR WAS REPLACED AND THE FAULT CLEARED. AS STATUS LEDS ARE TESTED PRIOR TO DISPATCH IT IS ASSUMED THAT THE TRANSISTOR AND COIN CELL HAD FAILED AFTER FINAL TEST AT HEARSTINE ON (B)(6) 2014. SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE OBSERVED SWITCHING ON AUTOMATICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 519143 | HEARTSINE SAMARITAN 350P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |