HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2017-00720
- Event Type
- Malfunction
- Date Received
- July 24, 2017
- Date of Event
- July 5, 2017
- Report Date
- October 3, 2017
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K014067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- PATIENT
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). AS THIS IS A PRE-2008 MODEL WHICH ONLY CONTAINS A USER ACCESSIBLE MEMORY LOG, NO INFORMATION CAN BE DETERMINED BETWEEN THE DEVICE PASSED 'OUT QAT' ON (B)(6) 2007 AND UPON RECEIPT AT HEARTSINE. THE DEVICE RECORDS MANUAL POWER UPS OF TEN MINUTES IN DURATION BETWEEN THE 21ST MAY 2014 AND THE 22ND MAY 2016. INVESTIGATION FOUND THE FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. MEMORY FULL PROMPT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 519133 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |