TRUERESULT
Report
- Report Number
- 1000113657-2017-01466
- Event Type
- Malfunction
- Date Received
- July 21, 2017
- Date of Event
- June 30, 2017
- Report Date
- July 21, 2017
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: (B)(4) USER'S TEST STRIP HAD POOR FILL.
CONSUMER REPORTED COMPLAINT FOR LO BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF LO. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120 - 140 MG/DL. CUSTOMER STATED HE ONLY TESTS THREE TIMES A WEEK. CUSTOMER STATED HE HAD NOT BEEN USING THE METER FOR A WHILE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6)2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/18/2018. CUSTOMER DID NOT REMEMBER THE OPEN VIAL DATE. TEST STRIPS ARE NOT IMPACTED BY RECALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: A 40MG/DL (B)(6)2017 03:24 PM FASTING: YES. A LO (B)(6)2017 02:17 PM FASTING: YES. A 42MG/DL (B)(6)2017 04:28 PM FASTING: YES. A 30MG/DL (B)(6)2017 04:28 PM FASTING: YES. A LO (B)(6)2017 04:28 PM FASTING :YES. MEMORY CONCERNS: CUSTOMER IS CONCERN WITH ALL THESE RESULTS. CUSTOMER STATES THAT THE METER IS GIVING RESULTS LOW RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 513703 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUERESULT | PS2465 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |