SEDLINE KIT
Report
- Report Number
- 2031172-2017-00731
- Event Type
- Malfunction
- Date Received
- July 21, 2017
- Date of Event
- June 27, 2017
- Manufacturer
- MASIMO - 40 PARKER
- Product Code
- GWQ
- PMA / PMN Number
- K151644
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL MANUFACTURING NARRATIVE (IF OTHER): THE RETURNED MODULE WAS EVALUATED. EXTERNAL VISUAL INSPECTION SHOWED A BROKEN OFF AND MISSING PIN 1 ON PATIENT CABLE. WHEN CONNECTED TO A MONITOR THE MONITOR DISPLAYED "HIGH IMPEDANCE DETECTED" AND NO WAVEFORM WAS DISPLAYED WITH TEST PLUG. THE MODULE WAS CONNECTED TO A MONITOR WITH A KNOWN GOOD PATIENT CABLE AND PASSED ALL FUNCTIONAL TESTS. THE CUSTOMER COMPLAINT WAS DUPLICATED FOR HIGH IMPEDANCE DUE TO THE BROKEN PIN ON THE PATIENT CABLE, THE MODULE PASSED ALL FUNCTIONAL TESTS AND THE CUSTOMER COMPLAINT REGARDING PSI VALUES WAS NOT DUPLICATED. INITIAL REPORTER ZIP CODE EXCEEDED MAXIMUM ALLOWABLE DIGITS, ZIP CODE IS AS FOLLOWS: (B)(6), CORRECTED DATA: CORRECTED FROM (B)(6). 510K# CORRECTED FROM K140188 TO K151644.
MULTIPLE ATTEMPTS FOR PRODUCT RETURN AND REQUESTS FOR ADDITIONAL INFORMATION WERE MADE. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF NEW INFORMATION IS OBTAINED OR THE PRODUCT IS RETURNED, A FOLLOW UP REPORT WILL BE SUBMITTED. (B)(6).
THE CUSTOMER REPORTED THE FOLLOWING ISSUE: "THERE HAPPENED THE ISSUE THAT CT IMPEDANCE INDICATOR BECAME BLUE. HOWEVER THE REPLACEMENT MODULE RESOLVED THE ISSUE. THERE WAS NO PROBLEM WITH THE SENSOR. THE IMPEDANCE READING WAS 10 TIMES COMPARED TO THE TYPICAL. ALSO, THERE WAS ANOTHER ISSUE. ALTHOUGH A PATIENT WAS UNDER SEDATION, THE ISSUE THAT PSI VALUE WAS VERY HIGH OCCURRED SEVERAL TIMES." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 512907 | SEDLINE KIT | ELECTROENCEPHALOGRAPH | GWQ | MASIMO - 40 PARKER | 25761 |