FDA Adverse Event
Injury
Summary report: N
STERIZONE VP4 STERILIZER
MDR report key: 6729960
·
Received July 20, 2017
Report
- Report Number
- 3004148947-2017-00001
- Event Type
- Injury
- Date Received
- July 20, 2017
- Date of Event
- June 20, 2017
- Report Date
- July 20, 2017
- Manufacturer
- TSO3 INC.
- Product Code
- PJJ
- PMA / PMN Number
- K141163
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION ON THE DEVICE WAS CONDUCTED ON SITE BY SERVICE TECHNICIAN. ALL TEST VERIFICATIONS WERE PERFORMED AND THE UNIT MET MANUFACTURER SPECIFICATIONS. THERE IS NO EVIDENCE THAT THIS EVENT WAS RELATED TO A DEVICE DEFECT OR MALFUNCTION. THE ISSUE IS ASSOCIATED WITH THE USE OF THE DEVICE IN TERMS OF NON-CONFORMING TO THE DEVICE'S INSTRUCTIONS AND INFORMATION PROVIDED BY THE DEVICE MANUFACTURER.
Description of Event or Problem · 1
IT WAS REPORTED THAT A STAFF MEMBER EXPERIENCED SLIGHT BURN/CHEMICAL IRRITATION ON FINGER FROM A SET THAT WAS IN STORAGE FOR THE OPERATING ROOM. THERE WAS NO REPORT OF A MEDICAL EVALUATION OR MEDICAL TREATMENT ASSOCIATED THE ALLEGED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 509530 | STERIZONE VP4 STERILIZER | TWO OR MORE STERILANT STERILIZER | PJJ | TSO3 INC. | 44000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |