TRIAGE PROFILER SOB PANEL 25T
Report
- Report Number
- 2027969-2017-00120
- Event Type
- Malfunction
- Date Received
- July 19, 2017
- Date of Event
- May 15, 2017
- Report Date
- June 22, 2017
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- DAP
- PMA / PMN Number
- K080269
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION CONCLUSION: THE CUSTOMER'S COMPLAINT WAS NOT REPLICATED DURING IN-HOUSE TESTING WITH RETAINED DEVICES OF LOT W62463B. NO ISSUES WITH D-DIMER RECOVERY WERE OBSERVED. MANUFACTURING BATCH RECORDS FOR THE LOT WERE REVIEWED AND FOUND THAT THE LOT MET RELEASE SPECIFICATION BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED.
INVESTIGATION PENDING. THIS EVENT IS BEING REPORTED AS PART NUMBER 97300EU IS SAME/SIMILAR TO 510 (K): K080269.
THE CUSTOMER REPORTED THE FOLLOWING EVENTS OCCURRING ON ONE PATIENT. THE CUSTOMER REPORTED BEING CONCERNED WITH THE D-DIMER RESULTS. ON (B)(6) 2017, THE PATIENT ARRIVED TO THE EMERGENCY DEPARTMENT WITH CHEST PAIN. AN INITIAL DRAW WAS PERFORMED AND BOTH WHOLE BLOOD AND PLASMA WERE COLLECTED IN EDTA: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 505622 | TRIAGE PROFILER SOB PANEL 25T | CARDIOPROFILER PANEL | DAP | ALERE SAN DIEGO, INC. | 97300EU | W62463B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TRIAGE METERPRO, PN 55070, SERIAL# (B)(4)| TRIAGE METERPRO, PN 55070, SERIAL# (B)(4)| TRIAGE METERPRO, PN 55070, SERIAL# (B)(4) |