FDA Adverse Event Malfunction Summary report: N

ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP

MDR report key: 6725213 · Received July 19, 2017

Report

Report Number
3006575795-2017-00244
Event Type
Malfunction
Date Received
July 19, 2017
Date of Event
July 3, 2017
Report Date
July 18, 2017
Manufacturer
ZYNO MEDICAL LLC
Product Code
FPA
UDI-DI
00814371020600
PMA / PMN Number
K132841
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE USER REPORTED LEAKING ISSUE WAS CONFIRMED. ZYNO MEDICAL HAS RECEIVED THE INVESTIGATION REPORT FROM THE CONTRACT SUPPLIER AMSINO INTERNATIONAL, INC ON 01/12/2018. THE ROOT CAUSE OF ISSUE IS FOUND TO BE THE DEFORMATION OF THE ROLLER OCCURED DURING THE ASSEMBLING PROCESS.

Additional Manufacturer Narrative · 1

THE MANUFACTURER IS CURRENTLY WAITING FOR THE ARRIVAL OF THE SET.

Description of Event or Problem · 1

THIS IS A FOLLOW-UP FOR THE INITIALLY FILED MDR (3006575795-2017-00244).

Description of Event or Problem · 1

ON (B)(6) THE USER REPORTED FLUID LEAKING ISSUE WITH ADMINISTRATION SETS. "WE HAVE HAD SEVERAL MORE CASES OF THIS SAME PROBLEM. WE KEPT ONE OF THE TUBING SETS. UNFORTUNATELY THE OTHER ONE WAS CONTAMINATED WITH HAZARDOUS DRUG AND WE HAD TO DISPOSE OF IT." THE NURSE USED 1000ML IVPB NORMAL SALINE, AND (B)(6)2017 WAS THE DATE THAT THE IVPB MADE IT TO THE NURSE. "TECHNICALLY THE PATIENT WAS NOT INVOLVED. THE MEDIATION WAS MADE AND SENT TO THE NURSE. BEFORE ADMINISTERING THE MEDICATION SHE NOTICED THAT FLUID WAS LEAKING INTO THE CHEMO BAG THAT CONTAINED THE IVPB. IT WAS NOTICED THAT THE CLAMP WAS CLOSED AND FLUID WAS STILL LEAKING. WE HAVE ALSO HAD THIS HAPPEN ON SEVERAL OCCASIONS WHEN WE SPIKE THE BAG BEFORE MEDICATION IS ADDED." NO PATIENT INJURY OR HARM OCCURRED FOR THIS CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
505591 ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP IV ADMINISTRATION SET FPA ZYNO MEDICAL LLC AA-80075 109233KS 00814371020600

Patients

Seq Age Sex Outcome Treatment
1