COBE CENTRYSYSTEM 3
Report
- Report Number
- 1713683-1997-00105
- Event Type
- Malfunction
- Date Received
- February 7, 1997
- Report Date
- January 9, 1997
- Manufacturer
- GAMBRO HEALTHCARE
- Product Code
- FII
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
THE LIP THAT HOLDS THE CAM IS EITHER WORN AND/OR BROKE OFF AND THE CAMS ARE FALLING OUT OR NOT OCCLUDING THE TUBING SET. THERE WERE NO RELATED CPF DEFECTS REPORTED FOR THIS MACHINE. TEST PROCEDURE FOLLOWED: QARA 146 SECTION 2.3, REVISION: G. THE FOUR-POSITION LINE CLAMP WAS NOT RETURNED FOR AN INVESTIGATION, BUT BASED ON PREVIOUS INVESTIGATIONS OF THIS PART (0896037C3, 0796207C3, 0796157C3, 0976160C3, AND 0796180C3) THE PART IS NOT NEEDED FOR THIS COMPLAINT INVESTIGATION. HOWEVER, THE ABOVE COMPLAINTS INDICATES THAT THE TABS ON THE WHITE CLAMP INSERT HOLDER WERE BROKEN OR STRETCHED, ALLOWING THE INSERTS TO FALL OUT OF THE HOLDER. IT WAS ALSO DETERMINED THAT THE RETURNED CLAMP'S STOP TABS WHICH HOLD THE CLAMP IN THE OCCLUDED POSITION WERE EITHER WORN OR BROKEN, SO THAT THE CLAMP NO LONGER STOPPED SECURELY IN THE CLOSED POSITION. THE ABOVE INFORMATION INDICATES THAT THE LINE CLAMP FAILURE DESCRIBED IN THIS COMPLAINT WAS GENERATED BY A BROKEN CLAMP ASSEMBLY. THIS ISSUE WILL CONTINUE TO BE TRENDED AS PART OF THE CO CORRECTIVE ACTION SYSTEM AND THE MANAGEMENT REVIEW TEAM. ANY FURTHER INVESTIGATIONS, ANALYSIS, AND/OR CORRECTIVE ACTIONS TAKEN ON THIS COMPLAINT ISSUE WILL BE DETERMINED BY AND DOCUMENTED IN THIS CORRECTIVE ACTION REVIEW PROCESS. SEE FAR #960027. FAILURE CODES: F. CODE SLNCL902W7. CUSTOMER CONTACTED: NO. SALES/SERVICE CONTACTED BY INVESTIGATOR? NO. TRAINING REQUIRED? NO.
DURING A DIALYSIS TREATMENT, A FOUR POSITION LINE CLAMP BROKE. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBE CENTRYSYSTEM 3 | DIALYSIS CONTROL UNIT | FII | GAMBRO HEALTHCARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |