HEAL COLLAR 3.5X3.5, 5MM
Report
- Report Number
- 0001038806-2017-00435
- Event Type
- Malfunction
- Date Received
- July 18, 2017
- Date of Event
- June 2, 2017
- Report Date
- October 10, 2017
- Manufacturer
- ZIMMER DENTAL
- Product Code
- NHA
- PMA / PMN Number
- PK111852
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
(B)(4). ONE ZIMMER DENTAL HEAL COLLAR WAS RETURNED FOR INSPECTION. VISUAL INSPECTION REVEALED WEAR ABOUT THE COLLAR AND THE THREADS. THE HEX HEAD WAS ALSO SLIGHTLY WORN, BUT FUNCTIONAL. PRODUCT DIMENSIONS WERE FOUND TO CONFORM WITH PRINT SPECIFICATIONS WHERE MEASURED AGAINST DRAWING REF ECO# 112456 REV A. THE IMPLANT THAT ALLEGEDLY WOULD NOT MATE WAS NOT RETURNED FOR INSPECTION, THEREFORE THE REPORTED EVENT COULD NOT BE VERIFIED. A DEVICE HISTORY REVIEW WAS PERFORMED AND NO RELATED NONCONFORMANCE¿S WERE NOTED. A COMPLAINT HISTORY SEARCH WAS PERFORMED USING OUR COMPLAINT HANDLING SYSTEM AND THERE WERE NO ADDITIONAL COMPLAINTS GENERATED AGAINST THIS LOT NUMBER. APPROPRIATE DOCUMENTATION WAS REVIEWED AND THE FOLLOWING INFORMATION WAS IDENTIFIED: INSTRUCTIONS FOR USE FOR HEALING COLLARS, HEALING SCREWS, SURGICAL COVER SCREWS AND TEMPORARY GINGIVAL CUFFS 9658 REV 1-01/15. HEALING COLLARS ¿ FOR USE WITH TAPERED SCREW-VENT®, SCREW-VENT® AND TRABECULAR METAL¿ IMPLANTS. SELECT THE APPROPRIATE SIZE HEALING COLLAR FOR THE PLATFORM OF THE IMPLANT AND WITH APPROPRIATE HEIGHT AND EMERGENCE PROFILE TO FIT THE EXISTING OR DESIRED TISSUE CONTOUR OF THE SITE. THE HEALING COLLARS ARE ANODIZED WITH COLOR-CODING ON THE LOWER PORTION TO INDICATE THE IMPLANT PLATFORM DIAMETER (FIGURE A). THE TOP SURFACE OF THE HEALING COLLAR IS ETCHED WITH THREE NUMBERS (FIGURE B) TO REFERENCE IMPLANT PLATFORM DIAMETER (LEFT), EMERGENCE PROFILE. DIAMETER (TOP RIGHT) AND CUFF HEIGHT (LOWER RIGHT). THE NUMBERS FOR IMPLANT PLATFORM AND EMERGENCE PROFILE SHOW ONLY THE INITIAL DIGIT OF THE RELATED MEASUREMENT. A SINGULAR CAUSE CANNOT BE DETERMINED. UDI NUMBER: (B)(4). THE DEVICE RETURN DATE IS (B)(6) 2017.
DEVICE HAS NOT YET BEEN RETURNED TO MANUFACTURE. PRODUCT NO RETURNED TO MANUFACTURE.
IT WAS REPORTED THAT HEALING COLLAR DOES NOT ASSEMBLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 502797 | HEAL COLLAR 3.5X3.5, 5MM | HEALING ABUTMENT | NHA | ZIMMER DENTAL | 63571118 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |