COMPREHENSIVE REVERSE SHOULDER CENTRAL SCREW
Report
- Report Number
- 0001825034-2017-04979
- Event Type
- Injury
- Date Received
- July 14, 2017
- Date of Event
- October 4, 2016
- Report Date
- February 2, 2018
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWS
- PMA / PMN Number
- PK132239
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE FOLLOWING REPORT IS SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE REPORTED EVENT IS CONFIRMED WITH REVIEW OF THE PROVIDED SURGICAL NOTES AS WELL AS THE RETURNED PRODUCTS. VISUAL EXAMINATION SHOWED GOUGES, NICKS AND EXCESSIVE WEAR. THIS VISUAL ABRASION CONFIRMS THE MIGRATION OF THE GLENOSPHERE, WHICH IS EVIDENCE OF DISASSOCIATION. AMONG THE RETURNED ITEMS WAS A SCREW NOT COMPATIBLE WITH THE OTHER IMPLANTS; WHILE THIS COULD HAVE BEEN A CONTRIBUTING FACTOR IT CANNOT BE CONFIRMED WHETHER THIS SCREW WAS USED IN THE PRIMARY PROCEDURE. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). MEDICAL PRODUCTS-BASEPLATE CATALOG#:010000589 LOT#:184750, LOCKING SCREW CATALOG#:180550 LOT#:808790, SCREW CATALOG#:180560 LOT#:119660, SCREW CATALOG#:180561 LOT#:224880, BEARING CATALOG#:XL-115363 LOT#:169800, STEM CATALOG#:113617 LOT#:909350, GLENOSPHERE CATALOG#:115316 LOT#:321870, TRAY CATALOG#:115370 LOT:910010, DRILL CATALOG#:405883 LOT#:562300, PIN CATALOG#:405800 LOT#:193340. THE DEVICE HAS BEEN RECEIVED BY ZIMMER BIOMET (B)(4) FOR EVALUATION. ONCE THE INVESTIGATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS:0001825034-2017-02775, 0001825034-2017-04973, 0001825034-2017-04976, 0001825034-2017-04977, 0001825034-2017-04978.
IT IS REPORTED THE PATIENT UNDERWENT A RIGHT TOTAL SHOULDER ARTHROPLASTY REVISION APPROXIMATELY FOUR MONTHS POST-IMPLANTATION DUE TO DISASSOCIATION OF THE TAPER FROM THE BASEPLATE. ADDITIONAL INFORMATION RECEIVED IN REVISION OPERATIVE NOTES, ALSO NOTES THE PATIENT WAS EXPERIENCING PAIN AND A GRINDING SENSATION PRIOR TO REVISION. DURING THE REVISION, SIGNIFICANT LINER EROSION WAS NOTED, AND THE BASEPLATE WAS FOUND TO BE LOOSE, WITH MARKED GLENOID BONE DEFICIENCY. ALL COMPONENTS, EXCEPT THE HUMERAL STEM, WERE REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 496246 | COMPREHENSIVE REVERSE SHOULDER CENTRAL SCREW | PROSTHESIS, SHOULDER | KWS | BIOMET ORTHOPEDICS | N/A | 399510 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| R | REFERENCE H10 |