SYMPHONY
Report
- Report Number
- 1000165971-2017-00557
- Event Type
- Injury
- Date Received
- July 12, 2017
- Date of Event
- June 21, 2017
- Report Date
- June 21, 2017
- Manufacturer
- SORIN CRM
- Product Code
- NVZ
- PMA / PMN Number
- P950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
PLEASE REFER TO THE ATTACHED ANALYSIS REPORT.
IT WAS REPORTED THAT TIME TO ERI WAS OVER-ESTIMATED. ON (B)(6) 2017 THE BATTERY IMPEDANCE WAS 5.75 KOHMS, THE MAGNET RATE WAS 93 MIN-1 AND THE TIME TO REPLACEMENT WAS DISPLAYED AS 16 MONTHS +/- 1 MONTH. ON (B)(6) 2017, LESS THAN 6 MONTHS LATER, THE TRANSTELEPHONIC MAGNET RATE READING WAS 81.6 MIN-1 INDICATING ERI COULD BE REACHED WITHIN WEEKS. THE RESULTS OF MONTHLY TTM CHECKS ARE AS FOLLOWS: (B)(6) 2017: 91.5 MIN-1, (B)(6) 2017: 91.5 MIN-1, (B)(6) 2017: 89 MIN-1, (B)(6) 2017: 88 MIN-1, (B)(6) 2017: 81.6 MIN-1. THE DEVICE WAS PROGRAMMED IN DDD MODE WITH RATE RESPONSE. A CHANNEL: 75% PACED, 2.5 V/0.5 MS, 453 OHMS AND V CHANNEL: 95% PACED, 2.5 V/0.5 MS, 5.22 OHMS.
IT WAS REPORTED THAT TIME TO ERI WAS OVER-ESTIMATED. ON (B)(6) 2017 THE BATTERY IMPEDANCE WAS 5.75 KOHMS, THE MAGNET RATE WAS 93 MIN-1 AND THE TIME TO REPLACEMENT WAS DISPLAYED AS 16 MONTHS +/- 1 MONTH. ON JUNE 21, 2017, LESS THAN 6 MONTHS LATER, THE TRANSTELEPHONIC MAGNET RATE READING WAS 81.6 MIN-1 INDICATING ERI COULD BE REACHED WITHIN WEEKS. THE RESULTS OF MONTHLY TTM CHECKS ARE AS FOLLOWS: (B)(6) 2017: 91.5 MIN-1, (B)(6) 2017: 91.5 MIN-1, (B)(6) 2017: 89 MIN-1, (B)(6) 2017: 88 MIN-1, (B)(6) 2017: 81.6 MIN-1. THE DEVICE WAS PROGRAMMED IN DDD MODE WITH RATE RESPONSE. A CHANNEL: 75% PACED, 2.5V/0.5 MS, 453 OHMS AND V CHANNEL: 95% PACED, 2.5V/0.5 MS, 5.22 OHMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 487195 | SYMPHONY | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | SORIN CRM | RHAPSODY + DR 2530 | M040715 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |