FDA Adverse Event Malfunction Summary report: N

LOCKING TOWER ENHANCED

MDR report key: 6705105 · Received July 12, 2017

Report

Report Number
3005180920-2017-00370
Event Type
Malfunction
Date Received
July 12, 2017
Date of Event
June 12, 2017
Report Date
November 15, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(6) 2017 THE (B)(4) PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEMS AND COMMENTED AS FOLLOWS: THE HINGES OF THE INSTRUMENTS GOT BLOCKED OR PRESENT EXCESSIVE FRICTION. POSSIBLY DUE TO LACK OF LUBRICATION AND/OR FOREIGN BODIES BETWEEN THE SLIDING SHAFTS OF THE INFERIOR PART OF THE INSTRUMENT, RESULTING IN EXCESSIVE FORCE ON THE HINGES THAT CONSEQUENTLY GOT DAMAGED AND/OR BLOCKED. THIS INSTRUMENT IS MAINLY USED FOR ROD REDUCTION INTO THE PEDICLE SCREW. THE SAME MANEUVER CAN BE ACHIEVED WITH ALTERNATIVE INSTRUMENT INCLUDED IN THE SET. BATCH REVIEW PERFORMED ON (B)(6) 2017: LOT 1651675: 46 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2016. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, THIS IS THE THIRD SIMILAR EVENT REPORTED ON ITEMS OF THE SAME LOT.

Additional Manufacturer Narrative · 1

INSTRUMENTATION WAS REPORTED AS AVAILABLE FOR INVESTIGATION BY THE INITIAL REPORTER, HOWEVER THE PIECE WAS NEVER SENT BACK TO MEDACTA INTERNATIONAL SA. THE COMPLAINT HAS BEEN CLOSED WITHOUT VISUAL INSPECTION OF THE ITEM. IN CASE OF ITEM RETURN, A FOLLOW-UP REPORT WILL BE SENT TO FDA. ON 18 SEPTEMBER 2017 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 25 SEPTEMBER 2017 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Description of Event or Problem · 1

DURING SURGERY, 6 LOCKING TOWERS JAMMED. THE SURGEON USED EXTRA INSTRUMENTATION TO COMPLETE THE SURGERY. THERE WAS A DELAY OF APPROXIMATELY 20-25 MINUTES. THE SURGERY WAS COMPLETED SUCCESSFULLY. INSTRUMENTATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
489370 LOCKING TOWER ENHANCED SURGICAL INSTRUMENT FOR SPINE LXH MEDACTA INTERNATIONAL SA

Patients

Seq Age Sex Outcome Treatment
1 Other