FDA Adverse Event Malfunction Summary report: N

COLIBRI II

MDR report key: 6701358 · Received July 11, 2017

Report

Report Number
8030965-2017-13559
Event Type
Malfunction
Date Received
July 11, 2017
Date of Event
June 14, 2017
Report Date
June 16, 2017
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. DURING REPAIR, IT WAS DETERMINED THAT THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER MAINTENANCE, WHICH IS USER ERROR/MISUSE/ABUSE. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MOTOR ON THE SMALL BATTERY DRIVE DEVICE MOTOR DID NOT FUNCTION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR: CHECK THE TRIGGERS AND ELECTRONIC CONTROL UNIT (ECU) FUNCTION, CHECK THE OFF/OSCION SWITCH MODE FUNCTION, CHECK THE OSCILLATION ANGLE, CHECK SWITCHING FUNCTION, CHECK THE FUNCTION WITH EPD-CONSOLE, CHECK COMPATIBILITY BETWEEN COLLBRLLSBD AND COLLBRI II /SBD II AND CHECK POWER WITH TEST BENCH, MLN.67.5 TO 110W. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK ANYMORE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
480590 COLIBRI II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1