SOFT PATH
Report
- Report Number
- 1058332-2006-00002
- Event Type
- Malfunction
- Date Received
- January 26, 2006
- Report Date
- December 6, 2001
- Manufacturer
- SCC SOFT COMPUTER
- Product Code
- LNX
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE CLIENT HAD MADE A DECISION TO USE THE SOFTPATH MODULE IN A NONSTANDARD MANNER. THE CLIENT'S STANDARD METHOD OF OPERATION IS TO REGISTER AND PROCESS A CASE FOR THE CG ACCESSION BASE AND THEN SIGN OUT THE CASE WITHOUT ENTERING ANY TYPE OF RESULTS. THEREFORE A FINAL REPORT HAS NO DIANGOSIS TEXT. LATER THE CLIENT CREATED A SUPPLEMENTAL REPORT TO ISSUE DIAGNOSTIC INFO. IMMEDIATELY AFTER CREATING THIS REPORT THE USER ACCESSES THE "CREATE NEW" SELECTION IN THIS OPTION FOR ANOTHER CASE. THE SYSTEM THEN CREATED A "FINAL" DOCUMENT FOR THE CURRENT CASE AND THE TEXT FROM THE PREVIOUSLY PRINTED CASE FOR ANOTHER PT IS SAVED FOR THIS PT. THIS ANOMALY OCCURS FOR TWO REASONS: 1. THE "FINAL REPORT" WAS NOT CREATED AS THE SYSTEM EXPECTED. 2. THE TEMPORARY DIRECTORY THAT IS CREATED DURING THE "PRINT PENDING" OPERATION IS NOT CLEARED UNTIL EXITING TO THE MAIN MENU. SUBSEQUENTLY THE PT RECEIVED AN ADT UPDATE TRANSACTION AND THIS EVENT TRIGGERED A RE-TRANSMITTAL OF REPORTS FROM SOFTPATH RESULT REPORTING INTERFACE. THE INCORRECT FINAL REPORT IS TRANSMITTED TO THE HIS AT THAT TIME AND THAT LEAD TO THE COMPLAINT OF THIS DESCRIBED SITUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SOFT PATH | LAB INFO SYSTEM | LNX | SCC SOFT COMPUTER | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |