FDA Adverse Event Malfunction Summary report: N

SOFT PATCH

MDR report key: 670101 · Received January 26, 2006

Report

Report Number
1058332-2006-00004
Event Type
Malfunction
Date Received
January 26, 2006
Report Date
September 17, 2004
Manufacturer
SCC SOFT COMPUTER
Product Code
LNX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TEXT FROM ONE CASE IS OVERWRITING ANOTHER. CASE RSC-24027 WAS ENTERED AND SAVED AND THEN CASE RSC-23945 WAS ENTERED AND SAVED. LATER REVIEW OF THE DOCUMENT OF 23945 REVEALED THAT THE TEXT WAS OVERWRITTEN BY CASE 24027. REVIEW OF THE SOFTWARE FAILURE REVEALED THAT THE TEMPORARY DIRECTORY WAS NOT CLEARED BEFORE PROCESSING EACH SUBSEQUENT CASE DURING ONE SESSION OF RESULTS ENTRY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOFT PATCH LAB INFO SYSTEM LNX SCC SOFT COMPUTER * *

Patients

Seq Age Sex Outcome Treatment
1 *