SYMPHONY
Report
- Report Number
- 1000165971-2017-00552
- Event Type
- Malfunction
- Date Received
- July 11, 2017
- Date of Event
- June 19, 2017
- Report Date
- June 19, 2017
- Manufacturer
- SORIN CRM
- Product Code
- NVY
- PMA / PMN Number
- P950029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- PHYSICIAN
Narratives
PLEASE REFER TO THE ATTACHED ANALYSIS REPORT.
REPORTEDLY, ON (B)(6) 2017, THE DEVICE WAS INTERROGATED DURING ROUTINE FOLLOW-UP. BATTERY LONGEVITY WAS 48 MONTHS AT BEGINNING OF SESSION (BATTERY IMPEDANCE WAS 3.09 KOHM AND MAGNET RATE WAS 95). NO HIGH OUTPUTS WERE PROGRAMMED. COMPARING TO PREVIOUS FOLLOW-UPS, LONGEVITY WAS INCREASED: 105 MONTHS ON (B)(6) 2015, 32 MONTHS ON (B)(6) 2015 AND 24 MONTHS (BATTERY IMPEDANCE 2, 7 KOHM AND 96 MAGNET RATE) ON (B)(6) 2016. AT THE END OF SESSION LONGEVITY WAS 52 MONTHS. PRELIMINARY ANALYSIS CONFIRMED THE REPORTED EVENT AND BASED ON AVAILABLE DATA SHOWED NORMAL BATTERY DEPLETION (AS OF (B)(6) 2017).
REPORTEDLY, ON (B)(6) 2017, THE DEVICE WAS INTERROGATED DURING ROUTINE FOLLOW-UP. BATTERY LONGEVITY WAS 48 MONTHS AT BEGINNING OF SESSION (BATTERY IMPEDANCE WAS 3.09 KOHM AND MAGNET RATE WAS 95). NO HIGH OUTPUTS WERE PROGRAMMED. COMPARING TO PREVIOUS FOLLOW-UPS, LONGEVITY WAS INCREASED: 105 MONTHS ON (B)(6) 2015, 32 MONTHS ON (B)(6) 2015 AND 24 MONTHS (BATTERY IMPEDANCE 2, 7 KOHM AND 96 MAGNET RATE) ON (B)(6) 2016. AT THE END OF SESSION LONGEVITY WAS 52 MONTHS. PRELIMINARY ANALYSIS CONFIRMED THE REPORTED EVENT AND BASED ON AVAILABLE DATA SHOWED NORMAL BATTERY DEPLETION (AS OF (B)(6) 2017).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484605 | SYMPHONY | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVY | SORIN CRM | SYMPHONY DR 2550 | S080131 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |