TECNIS 1-PIECE
Report
- Report Number
- 3011852734-2017-00008
- Event Type
- Malfunction
- Date Received
- July 10, 2017
- Date of Event
- June 8, 2017
- Report Date
- November 6, 2017
- Manufacturer
- ABBOTT MEDICAL OPTICS
- Product Code
- HQL
- UDI-DI
- 05050474531772
- PMA / PMN Number
- P980040
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.
PRODUCT TESTING COULD NOT BE PERFORMED BECAUSE THE PRODUCT WAS NOT RETURNED. THE REPORTED COMPLAINT WAS NOT VERIFIED. THE MANUFACTURING RECORDS FOR THE PRODUCT WERE REVIEWED. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATIONS. A SEARCH REVEALED THAT NO ADDITIONAL COMPLAINTS FOR THIS ORDER NUMBER HAVE BEEN RECEIVED. THE DIRECTIONS FOR USE (DFU) WERE REVIEWED. THE DFU PROVIDE THE CUSTOMER WITH PROPER USAGE INSTRUCTIONS AND GUIDELINES. AS A RESULT OF THE INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS SCRATCHED DURING INSERTION INTO THE EYE. THERE WAS PATIENT CONTACT. NO ADDITIONAL INFORMATION WAS PROVIDED TO ABBOTT MEDICAL OPTICS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 477677 | TECNIS 1-PIECE | MONOFOCAL IOLS | HQL | ABBOTT MEDICAL OPTICS | ZCB00 | 05050474531772 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |