FDA Adverse Event Malfunction Summary report: N

SOFTLAB

MDR report key: 669937 · Received January 26, 2006

Report

Report Number
1058332-2006-00010
Event Type
Malfunction
Date Received
January 26, 2006
Report Date
November 5, 2004
Manufacturer
SCC SOFT COMPUTER
Product Code
LNX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

THE CLIENT REQUESTED ASSISTANCE IN CREATING A NEW REPORT LAYOUT. DURING THE COURSE OF TESTING, THE CLIENT NOTICED THAT THE INCORRECT REFERENCE RANGES WERE BEING ASSIGNED TO TESTS. DURING SCC'S INVESTIGATION IT WAS DISCOVERED THAT AN OBSOLETE VARIABLE RELATED TO REFERENCE RANGES WAS COPIED INTO THE CLIENT'S NEWLY CUSTOMIZED REPORT LAYOUT. THIS PROBLEM CAN LEAD TO INCORRECT REFERENCE RANGES BEING APPLIED TO THE PT REPORT. THE PROBLEM WAS NOT WITH THE SOFTWARE, BUT WITH THE MANUAL COPYING OF AN OLDER REPORT LAYOUT VARIABLE INTO THE NEWLY CUSTOMIZED LAYOUT WHICH OVERWROTE THE CORRECT VARIABLE WITH THE OBSOLETE VARIABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOFTLAB LAB INFO SYSTEM LNX SCC SOFT COMPUTER * *

Patients

Seq Age Sex Outcome Treatment
1 *