STEALTHSTATION S7 SYSTEM
Report
- Report Number
- 1723170-2017-02799
- Event Type
- Injury
- Date Received
- July 7, 2017
- Date of Event
- February 29, 2016
- Report Date
- August 25, 2017
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
CORRECTION: MULTIPLE ATTEMPTS WERE MADE TO THE AUTHOR TO OBTAIN ADDITIONAL INFORMATION ON SYSTEM SERIAL NUMBER. THE MEDTRONIC REPRESENTATIVE OF THE SITE SUSPECTS THE PROVIDED INFORMATION TO BE THE SYSTEM USED IN THE JOURNAL ARTICLE. BRAND NAME, COMMON DEVICE NAME AND PROCODE NOW PROVIDED. DEVICE UDI, CATALOG NUMBER, AND 510K NOW CORRECTED. ADDITIONAL INFORMATION: DEVICE SERIAL NUMBER NOW PROVIDED. ADDITIONAL INFORMATION: DEVICE MANUFACTURING DATE NOW PROVIDED.
SPECIFIC PATIENT IDENTIFIER NOT AVAILABLE FROM THE JOURNAL ARTICLE AUTHORS. PATIENT SEX AND WEIGHT NOT AVAILABLE FROM THE JOURNAL ARTICLE AUTHORS. EVENT DATE IS APPROXIMATED. DATE PROVIDED IS USED AS THE PATIENTS UNDERWENT SURGERY THROUGH (B)(6) 2016. CITATION: JIN, M., LIU, Z., QIU, Y., YAN, H., HAN, X. & ZHU, Z. INCIDENCE AND RISK FACTORS FOR THE MISPLACEMENT OF PEDICLE SCREWS IN SCOLIOSIS SURGERY ASSISTED BY O-ARM NAVIGATION¿ANALYSIS OF A LARGE SERIES OF ONE THOUSAND, ONE HUNDRED AND FORTY FIVE SCREWS. INTERNATIONAL ORTHOPAEDICS (SICOT) (2017) 41:773¿780. DOI 10.1007/S00264-016-3353-6. BRAND NAME, COMMON DEVICE NAME AND PROCODE NOT PROVIDED IN THE JOURNAL ARTICLE. THE ARTICLE MENTIONS A NAVIGATION SYSTEM (MODEL NOT SPECIFIED). FURTHER INFORMATION UNAVAILABLE. DEVICE SERIAL NUMBER, UNAVAILABLE. DEVICE UDI NOT PROVIDED AS THIS PRODUCT IS NO LONGER MANUFACTURED. DEVICE MANUFACTURING DATE IS DEPENDENT ON SERIAL NUMBER, THEREFORE, UNAVAILABLE. THE JOURNAL ARTICLE REPORTS THAT UNIVARIATE ANALYSIS DEMONSTRATED THAT RISK FACTORS ASSOCIATED WITH SCREW MALPOSITIONING INCLUDED AETIOLOGY, THE INSTRUMENTED LEVEL, GAP DIFFERENCE, AND CONCAVE PEDICLE. PATHOLOGICAL CONDITION (CS, NF-1, ETC.) MIDDLE-THORACIC SPINE, GAP DIFFERENCE OVER THREE LEVELS, AND CONCAVE PEDICLES WERE RISK FACTORS FOR PREDICTING PEDICLE SCREW MISPLACEMENT (TABLE 4). AMONG THESE, GAP DIFFERENCE (OVER 3 LEVELS) RANKS THE HIGHEST OF ALL. THE STUDY FOUND THAT PATIENTS WITH NF-1 AND CS WERE AT HIGH RISK OF PEDICLE SCREW MISPLACEMENT. THE REASONS MIGHT BE AS FOLLOWS: FIRST, THE DYSPLASIA STRUCTURES IN THESE CASES MIGHT RESULT IN IMPROPER FIXATION OF THE TRACKER TO THE IMPLANT, LEADING TO A DIFFERENCE IN ANGULATION BETWEEN THE VIRTUAL AND THE INTRAOPERATIVE IMAGES. SECOND, THE SEVERE DISTORTED AND ROTATED PEDICLES IN THESE CASES POSED GREAT CHALLENGE FOR FINDING THE OPTIMAL TRAJECTORY OF SCREWS. ACCURATE PLACEMENT OF PEDICLE SCREWS IN SCOLIOSIS REMAINS A CHALLENGE, ESPECIALLY IN SEVERE CASES AS THE PEDICLES ARE THINNER, MORE SCLEROTIC AND DYSPLASTIC THAN THOSE OF MILD SCOLIOSIS. THE JOURNAL ARTICLE ALSO REPORTS THAT, ALTHOUGH THE NAVIGATION SYSTEM IS EXTRAORDINARY CONVENIENT, SEVERAL TIPS ARE NECESSARY FOR THE FUTURE SURGEONS TO SHORTEN THE LEARNING CURVE AND AVOID THE PITFALLS OF THIS TECHNIQUE: IN CASES WHERE THE PEDICLES ARE DYSPLASIA SUCH AS NF-1 AND CS, THE PERFORATION RATE COULD BE STILL HIGH DESPITE OF NAVIGATION; EVEN IF UNDER GREAT VISION, INSTRUMENTATION OF THORACIC SPINE ESPECIALLY UPPER THORACIC SPINE COULD BE DANGEROUS; IT¿S BETTER TO LOCATE THE TRACKER WITHIN THE APICAL REGION OR AROUND INSTRUMENTED LEVELS AS CLOSE AS POSSIBLE; INSERTING THE PEDICLE SCREWS FROM DISTALLY TO PROXIMALLY TO THE TRACKER COULD REDUCE THE INCIDENCE OF TRACKER LOOSENING AND AVOID ACCIDENTAL RADIATION EXPOSURE DUE TO RE-SCAN. PEDICLE PERFORATIONS ARE KNOWN INHERENT RISKS TO ANY SPINAL SURGICAL PROCEDURE. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INFORMATION PROVIDED IN THE JOURNAL ARTICLE OR FROM THE AUTHORS. NO REQUEST FOR SERVICE HAVE BEEN RECEIVED FROM THE CUSTOMER REGARDING THESE EVENTS. NO PARTS HAVE BEEN REPLACED OR RETURNED TO THE MANUFACTURER FOR EVALUATION.
THE JOURNAL ARTICLE WAS FORWARDED BY A MEDTRONIC REPRESENTATIVE. USE OF THE USE OF THE NAVIGATION PLUS SYSTEM AND IMAGING SYSTEM WAS REPORTED FOR SCOLIOSIS TREATMENT WITH PEDICLE SCREW PLACEMENT BETWEEN SEPTEMBER 2013 AND FEBRUARY 2016. THIS IS MEDICAL DEVICE REPORT (MDR) REPORT 2 OF 2. SEE 1723170-2017-02798. THE PATIENTS WERE CATEGORIZED INTO FOUR GROUPS ACCORDING TO PATHOLOGICAL CONDITIONS. ALL SURGERIES WERE DONE BY TWO SENIOR SURGEONS WITH AT LEAST TEN YEARS OF OPERATIVE EXPERIENCE IN SPINE SURGERY AND VERY FAMILIAR WITH IMAGING AND NAVIGATION SYSTEM. A FIRST INTRA-OPERATIVE 3D SCANNING WAS ACQUIRED INCLUDING FOUR TO SIX VERTEBRAL LEVELS, DEPENDING ON THE SIZE OF VERTEBRAE. ONCE ALL SCREWS WERE IN PLACE, THEIR POSITIONS WERE NOT EVALUATED WITH A SECOND INTRA-OPERATIVE 3D SCAN WITH THE IMAGING SYSTEM IN ORDER TO REDUCE THE EXTRA RADIATION EXPOSURE FOR THE PATIENT AS WELL AS FOR THE STAFF. PEDICLE SCREW VIOLATIONS WERE CATEGORIZED INTO FOUR GRADES: GRADE 0, SCREWS WERE COMPLETELY WITHIN THE PEDICLE; GRADE 1, PENETRATION LESS THAN 2 MM; GRADE 2, PENETRATION BETWEEN 2 AND 4 MM; AND GRADE 3, PENETRATION OVER 4 MM. ADDITIONALLY, GRADE 0 AND 1 PERFORATIONS WERE CONSIDERED SATISFACTORY, WHILE GRADE 2 AND 3 WERE REGARDED AS PERFORATION (FIG. 1). A (B)(6)-YEAR-OLD PATIENT WITH ADOLESCENT IDIOPATHIC SCOLIOSIS (AIS) UNDERWENT POSTERIOR INSTRUMENTATION FROM T3 TO L1. PRE-OPERATIVE POSTEROANTERIOR (PA) AND LATERAL X-RAY SHOWED MAIN THORACIC (MT) AND THORACIC KYPHOSIS (TK) COBB ANGLES WERE 80 AND 36¿RESPECTIVELY (FIG. 3). THE PEDICLE SCREWS WERE INSERTED FROM T5-T10 BY USING NAVIGATION SYSTEM, AND SIGNIFICANT PERFORATIONS WERE NOTED AT T5, T6, AND T8 ON POST-OPERATIVE CT IMAGES. HOWEVER, POST-OPERATIVE PA AND LATERAL X-RAY FILMS SHOWED SATISFYING CORONAL AND SAGITTAL CORRECTION. THE ARTICLE CONCLUDES, THAT THE STUDY HAS SHOWN THAT 90.2% OF PEDICLE SCREWS COULD BE PLACED ACCURATELY AND SAFELY IN SEVERE SCOLIOSIS SURGERIES WHILE USING THE INTRA-OPERATIVE NAVIGATION SYSTEM. HOWEVER, PATIENTS WITH CS AND NF-1 WERE AT HIGH RISK OF PEDICLE SCREW MISPLACEMENT, MOST OF WHICH WERE LOCATED AT MIDDLE THORACIC SPINE AND CONCAVE PEDICLES. MOREOVER, IT IS SUGGESTED TO PLACE THE TRACKER AS CLOSE AS POSSIBLE TO THE INSTRUMENTED LEVEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476264 | STEALTHSTATION S7 SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. (LOUISVILLE) | S7 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR | Other |