TRUERESULT
Report
- Report Number
- 1000113657-2017-01236
- Event Type
- Malfunction
- Date Received
- July 6, 2017
- Date of Event
- January 25, 2016
- Report Date
- July 6, 2017
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- OTHER
Narratives
INTERNAL REPORT # (B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE. BASED ON REVIEW WITH FDA ON 06/21/2017, WE VERIFIED THE AGENCY'S INTERPRETATION OF THE ACT AND SUBMIT THIS MEDWATCH REPORT BASED ON THE NATURE OF THE COMPLAINT FOR THIS RECALLED LOT. RECALL #FDA-ASSIGNED RECALL EVENT ID (B)(4).
CONSUMER REPORTED COMPLAINT FOR LOW RESULTS. CUSTOMER REPORTED NO SYMPTOMS, AND DOES NOT NEED MEDICAL ATTENTION. STRIP EXPIRATION DATE IS 06/30/2018 AND STRIP OPEN DATE IS (B)(6) 2016. STRIP STORAGE IS CORRECT. BACK TO BACK BLOOD TEST PERFORMED NON-FASTING AND RESULTS ARE 200MG/DL AND 190MG/DL REVIEWED METER MEMORY (DATE / TIME NOT SET) THE 176MG/DL (B)(6) 2016 08:47:00 AM FASTING:YES, THE 120MG/DL (B)(6) 2016 07:00:00 PM FASTING:YES, THE 171MG/DL (B)(6) 2015 10:43:00 PM FASTING:YES, THE 188MG/DL (B)(6) 2016 08:51:00 AM FASTING:YES, THE 143MG/DL (B)(6) 2015 06:22:00 AM FASTING:YES. CUSTOMER CALLING CONCERNED HIS METER IS READING LOWER THAN HIS DOCTOR'S METER. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST ON (B)(6) 2016 AT 09:05 AM FASTING AND RECEIVED A RESULT OF 192 MG/DL AND AT THE DOCTOR'S OFFICE USING THE FACILITY METER CUSTOMER RECEIVED A RESULT OF 255 MG /DL . CUSTOMER STATES HIS NORMAL GLUCOSE RANGE IS 132-165 MG /DL FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471825 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUERESULT | PS2383 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |