FDA Adverse Event Malfunction Summary report: N

REDUCT FORCEPS POINTS NARROW

MDR report key: 6682175 · Received June 30, 2017

Report

Report Number
0001825034-2017-04479
Event Type
Malfunction
Date Received
June 30, 2017
Date of Event
March 2, 2017
Report Date
June 29, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
HYA
PMA / PMN Number
PEXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-02428, 04479. COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. VISUAL REVIEW OF THE TWO (2) DEVICES RETURNED IDENTIFIED THAT ONE PRONG ON EACH OF THE FORCEPS IS FRACTURED. MATERIAL ANALYSIS STATED THE FOLLOWING: "THE CLAMPS HAVE SCRAPES AND SCRATCHES BUT APPEARED AS DESIGNED EXCEPT FOR NEARLY IDENTICAL FRACTURED TIPS. BOTH FRACTURES OCCURRED AT THE MOST DISTAL NOTCH NEARLY PERPENDICULAR TO THE TIP AXIS. FRACTURE ARTIFACTS SUGGEST A BENDING OVERLOAD FRACTURE MAY HAVE BEEN INITIATED. THE POSSIBLE DIRECTIONS OF THE FRACTURES SUGGEST THEY BROKE DURING CLAMPING. AN XRF SCAN CONFIRMED THE MATERIAL AS 410/420 STAINLESS STEEL. A FIELD PHOTO SHOWS THE CLAMPS IN THE OR WITH ONE OF THE FRACTURED TIPS, WHICH WAS NOT RETURNED." DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION(S) IS/ARE REQUIRED. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT IS REPORTED THAT THE TIP OF TWO FORCEPS FRACTURED DURING A PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462294 REDUCT FORCEPS POINTS NARROW SURGICAL INSTRUMENT, MANUAL HYA BIOMET ORTHOPEDICS N/A 0316GH

Patients

Seq Age Sex Outcome Treatment
1