MPACT CANCELLOUS BONE SCREW, FLAT HEAD Ø 6,5 L 50
Report
- Report Number
- 3005180920-2017-00350
- Event Type
- Malfunction
- Date Received
- June 30, 2017
- Date of Event
- June 2, 2017
- Report Date
- July 27, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LPH
- PMA / PMN Number
- K132879
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
ON 04 JULY 2017 THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEM AND COMMENTED AS FOLLOWS: THE PIECE WAS ANALYZED. THE SCREW IS BENT IN THE PART CLOSED TO THE SCREW HEAD. IT IS NOT CLEAR HOW IT WAS HAPPENED AND WE CAN NOT DETERMINATE WITH CERTAINTY THE ROOT CAUSE OF THE EVENT, BUT WE CAN ASSUME THAT IT IS DUE TO AN ERRONEOUS USE DURING SURGERY: IN ORDER TO CORRECT THE DIRECTION OF INSERTION, THE SURGEON WOULD HAVE APPLIED A FORCE IN FLEXION TO PROVIDE THE BENDING OF THE ACETABULAR SCREW.
ADDITIONAL INFORMATION RECEIVED ON 09 JUNE 2017 AND INCLUDES: IF THE SURGEON CONTINUED TO TIGHTEN THE SCREW, HE WOULD HAVE BROKEN IT. THE SURGEON USED THE UNIVERSAL HANDLE. BATCH REVIEW PERFORMED ON 26 JUNE 2017. LOT 132489: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11 DECEMBER 2013. EXPIRATION DATE: 2018-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT YET RECEIVED.
WHEN THE SURGEON WAS IMPLANTING THE SCREW, THE SCREW BENT. THE SURGEON USED A SMALLER SCREW TO COMPLETE THE SURGERY. THERE WAS A SHORT DELAY OF LESS THAN 2 MINUTES. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS ARE NOT AVAILABLE; IMPLANT IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460885 | MPACT CANCELLOUS BONE SCREW, FLAT HEAD Ø 6,5 L 50 | CANCELLOUS BONE SCREW | LPH | MEDACTA INTERNATIONAL SA | 132489 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |