FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 6680156 · Received June 30, 2017

Report

Report Number
3004123209-2017-00622
Event Type
Malfunction
Date Received
June 30, 2017
Date of Event
June 14, 2017
Report Date
August 23, 2017
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z00124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). ROUTINE TESTING CONFIRM THAT THE PAD-PAK WAS FIRST WAS INSTALLED BY THE CUSTOMER ON THE 7TH JUNE 2010. THE INVESTIGATION WAS UNABLE TO REPLICATE THE REPORTED FAULT. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE OBSERVED SWITCHING ON AUTOMATICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463741 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM

Patients

Seq Age Sex Outcome Treatment
1