FDA Adverse Event Malfunction Summary report: N

GLUCOMETER ELITE 4 DATE

MDR report key: 667865 · Received January 20, 2006

Report

Report Number
1826988-2006-00028
Event Type
Malfunction
Date Received
January 20, 2006
Date of Event
January 10, 2006
Report Date
January 12, 2006
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

CUSTOMER'S DAUGHTER STATES THAT HER FATHER'S READINGS HAVE BEEN ABNORMAL THIS WEEK--BOTH HIGHER AND LOWER. HIS NORMAL READINGS ARE 100-140MG/DL. HIS READINGS HAVE BEEN 184, 199, 233, 420, 88, 213. CUSTOMER SERVICE HAD THE CALLER CHECK THE MEMORY AND FOUND RESULTS OF 14.8, 18.4, 19.9, 23.3, 4.9, 17.6, 15.7 MMOL/L. THE CUSTOMER HDI NOT NOTICE THE DECIMAL POINT IN THE READINGS. CUSTOMER SERVICE HELPED THE CALLER SET THE METER BACK TO MG/DL. RESULTS IN MG/DL WERE 266, 332, 359, 213, 420, 88, 316, 283, 243MG/DL. CHECK TEST SHOWED THE METER WAS WORKING CORRECTLY. METER WAS CODED CORRECTLY TO F-5. CUSTOMER SERVICE DISCUSSED CORRECT TECHNIQUES. CUSTOMER PERFORMED TWO BLOOD TESTS AND VERIFIED PROPER FILL--RESULTS WERE 379 AND 390MG/DL. CUSTOMER SERVICE ADVISED THE CALLER TO CONTACT A DOCTOR WITH RESULTS 250+ OR -50MG/DL. CUSTOMER SERVICE ADVISED THE CUSTOMER TO RETURN THE METER AND WILL REPLACE WITH ANOTHER ELITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GLUCOMETER ELITE 4 DATE BLOOD GLUCOSE METER NBW BAYER HEALTHCARE, LLC 3901B NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN