CD1700 CS
Report
- Report Number
- 2919069-2006-00005
- Event Type
- Other
- Date Received
- January 18, 2006
- Date of Event
- December 22, 2005
- Report Date
- January 17, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
INTERNAL IDENTIFIERS (S): 060/1-163016463-11 PATIENT CODE: 2199 (PATIENT, NO CONSEQUENCES TO). DEVICE CODE: 2456 (INACCURATE TEST RESULTS) THE CUSTOMER TECHNICAL ADVOCATE SUGGESTED THAT THE CUSTOMER CLEAN THE CLOSED MODE OPERATION ASSEMBLY WITH BLEACH, FOLLOWED BY A DISTILLED WATER FLUSH. THE CUSTOMER ALSO REPLACED THE CLOSED MODE ASSEMBLY PUMP TUBING AND RAN SUCCESSFUL BACKGROUND COUNTS WITH CONTROLS. THE CUSTOMER ALSO PERFORMED A PRECISION RUN WITH 6 REPLICATES AND ALL RESULTS REPRODUCED WITHIN EXPECTED SPECIFICATIONS. THE CUSTOMER REQUESTED NO FURTHER ASSISTANCE. NO FURTHER ISSUES IDENTIFIED. THE CELL-DYN 1700 SYSTEM OPERATOR'S MANUAL, REVISION D: SECTION 10: TROUBLESHOOTING AND DIAGNOSTICS: TROUBLESHOOTING GUIDE: ABNORMAL OR ERRATIC HGB, MCH AND/OR MCHC RESULTS: PG 10-13, PROVIDES INSTRUCTIONS IN REGARDS TO THIS ISSUE ON CLEANING AND THE REPLACEMENT OF ASPIRATION TUBING. THIS INVESTIGATION DEMONSTRATED THAT THE CELL-DYN 1700 CS ANALYZER IS PERFORMING WITHIN ITS INTENDED USE, LABEL CLAIMS AND SPECIFICATIONS; NO DEFICIENCY RELATED TO THE PERFORMANCE OF THE DEVICE WAS IDENTIFIED. THIS IS A FINAL REPORT.
THE CELL-DYN 1700 CS GENERATED VERY LOW RESULTS FOR ALL PARAMETERS ON TWO DONOR SAMPLES. THE RESULTS WERE FLAGGED AS "L" OR "LL" INDICATING THAT THE RESULTS WERE OUTSIDE OF THE LAB'S ESTABLISHED REFERENCE RANGES. THE CUSTOMER PERFORMED AN AUTO-CLEAN PROCEDURE ON THE ANALYZER AND RAN CONTROLS, WHICH TESTED WITHIN SPECIFICATION. HOWEVER, PATIENT #2 GENERATED AN INITIAL HEMOGLOBIN RESULT OF 3.2 G/DL THAT RETESTED AT 14.4 G/DL WITH ALL OTHER PARAMETERS ALSO TESTING WITHIN THE LAB'S NORMAL REFERENCE RANGE. THE CUSTOMER SUGGESTED A NUMBER OF TROUBLESHOOTING PROCEDURES FOR THE CUSTOMER TO TRY. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CD1700 CS | AUTOMATED HEMATOLOGY ANALYZER | GKL | ABBOTT DIAGNOSTICS DIVISION/CELLDYN | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |