FDA Adverse Event Malfunction Summary report: N

FGA 450

MDR report key: 6674591 · Received June 28, 2017

Report

Report Number
3007802293-2017-00093
Event Type
Malfunction
Date Received
June 28, 2017
Date of Event
June 1, 2017
Report Date
June 28, 2017
Manufacturer
PRISM MEDICAL
Product Code
FSA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

OUTCOME OF INVESTIGATION PENDING. A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Description of Event or Problem · 1

THE FACILITY SHARED THE RESIDENT WAS BEING TRANSFERRED WITH ASSISTANCE OF TWO PEOPLE FROM HER BED TO HER WHEEL CHAIR USING A FGA 450 LIFT WITH A QUICK RELEASE EASY SWIVEL CARRY BAR AND A SIZE SMALL UNIVERSAL SLING. AS THE LIFT WAS BEING MOVED TOWARDS THE CHAIR THE SLING MIDDLE LOOP MIGRATED OFF THE LIFT HOOK ON THE RIGHT SIDE. THE RESIDENT'S UPPER BODY SLID TO THE RIGHT, THROUGH THE SIDE OF SLING, BETWEEN THE TWO DISTAL LOOPS. SHE FELL ONTO THE FLOOR. THE FACILITY REMOVED THE LIFT FROM SERVICE TO INVESTIGATE. THE RESIDENT RECEIVED MEDICAL TREATMENT. SHE HAD A HIP FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456694 FGA 450 FGA 450 LIFT FSA PRISM MEDICAL 280411

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization