FDA Adverse Event
Malfunction
Summary report: N
FGA 450
MDR report key: 6674591
·
Received June 28, 2017
Report
- Report Number
- 3007802293-2017-00093
- Event Type
- Malfunction
- Date Received
- June 28, 2017
- Date of Event
- June 1, 2017
- Report Date
- June 28, 2017
- Manufacturer
- PRISM MEDICAL
- Product Code
- FSA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
OUTCOME OF INVESTIGATION PENDING. A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION.
Description of Event or Problem · 1
THE FACILITY SHARED THE RESIDENT WAS BEING TRANSFERRED WITH ASSISTANCE OF TWO PEOPLE FROM HER BED TO HER WHEEL CHAIR USING A FGA 450 LIFT WITH A QUICK RELEASE EASY SWIVEL CARRY BAR AND A SIZE SMALL UNIVERSAL SLING. AS THE LIFT WAS BEING MOVED TOWARDS THE CHAIR THE SLING MIDDLE LOOP MIGRATED OFF THE LIFT HOOK ON THE RIGHT SIDE. THE RESIDENT'S UPPER BODY SLID TO THE RIGHT, THROUGH THE SIDE OF SLING, BETWEEN THE TWO DISTAL LOOPS. SHE FELL ONTO THE FLOOR. THE FACILITY REMOVED THE LIFT FROM SERVICE TO INVESTIGATE. THE RESIDENT RECEIVED MEDICAL TREATMENT. SHE HAD A HIP FRACTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456694 | FGA 450 | FGA 450 LIFT | FSA | PRISM MEDICAL | 280411 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Hospitalization |