SINGLE USE PAPER FILTER WITH INDICATOR
Report
- Report Number
- 9610612-2017-00345
- Event Type
- Malfunction
- Date Received
- June 28, 2017
- Date of Event
- June 2, 2017
- Report Date
- August 7, 2017
- Manufacturer
- AESCULAP AG
- Product Code
- OWP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.
(B)(6). IT WAS REPORTED THAT PARTICLES WAS FOUND INSIDE THE CONTAINERS THAT HAVE BEEN STERILIZED AS LONG AS TWO TO SIX MONTHS AGO. THE PARTICLES IN THE CONTAINERS ARE DARKER IN COLOR AND THE PARTICLES ARE SOLID. THE LIDS ALSO HAD SOME DUST ON THEM IN THE CREVICES.
INVESTIGATION: THE INVESTIGATION WAS CARRIED OUT VISUALLY. THE DUST IN QUESTION IS NOT AVAILABLE FOR INVESTIGATION, ONLY PICTURES WERE PROVIDED. THE SINGLE USE PAPER FILTER AND FILTER HOLDERS ARE USED, BUT NO DAMAGE CAN BE FOUND. THE CONTAINER BOTTOMS AND LIDS ARE ALSO IN USED CONDITION, SEVERAL SCRATCHES ON THE SURFACE, BUT NO SIGNIFICANT DAMAGES CAN BE FOUND. CONCLUSION AND ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF OUR INVESTIGATION, THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY USER RELATED. RATIONAL: THE PROVIDED PRODUCTS ARE ALL IN GOOD CONDITION, THUS WE EXCLUDE A MATERIAL OR MANUFACTURING RELATED ERROR. FURTHERMORE IT IS VERY UNLIKELY THAT SUCH A LARGE AMOUNT OF DUST CAN GET INTO A PROPERLY WORKING CONTAINER SYSTEM. IT IS CONCEIVABLE THAT SUCH AN AMOUNT OF IMPURITIES CAN GET INTO A CONTAINER, IF THE USER HAS FORGOTTEN TO INSTALL A PAPER FILTER BEFORE THE STERILIZATION PROCESS. NO CAPA IS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 454951 | SINGLE USE PAPER FILTER WITH INDICATOR | DISPOSABLES CONTAINER SYSTEM | OWP | AESCULAP AG | JK095 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |