FDA Adverse Event
Injury
Summary report: N
TELEDYNE WATER PIK
MDR report key: 66721
·
Received January 30, 1997
Report
- Report Number
- 1712259-1997-00002
- Event Type
- Injury
- Date Received
- January 30, 1997
- Date of Event
- December 31, 1996
- Report Date
- January 30, 1997
- Manufacturer
- TELEDYNE WATER PIK
- Product Code
- EFS
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED UNIT WAS VERY DIRTY. ALL PERFORMANCE VALUES WERE IN SPEC. EXCEPT FOR HIGH PRESSURE THAT WAS 68PSI WHEN COMPARED TO THE SPECIFICATION - 75PSI MINIMUM.
Description of Event or Problem · 1
CUSTOMER CALLED TO STATE THAT THE ORAL IRRIGATOR FORCED FOOD DEBRIS UNDER HER DENTAL WORK CAUSING DAMAGE TO THE BONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TELEDYNE WATER PIK | ORAL IRRIGATOR | EFS | TELEDYNE WATER PIK | WP-20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Disability |