FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 6670330 · Received June 27, 2017

Report

Report Number
3004209178-2017-57805
Event Type
Malfunction
Date Received
June 27, 2017
Date of Event
February 21, 2017
Report Date
June 2, 2017
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD BLANK DISPLAY DUE TO MISSING BATTERY TUBE SPRING. UNABLE TO PERFORM IDLE CURRENT TEST, RUN CURRENT TEST, SELF-TEST, OFF NO POWER TEST, UNEXPECTED ERROR TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST DUE TO BLANK DISPLAY. PUMP WAS RECEIVED WITH CORRODED BATTERY TUBE AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS DAMAGED. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE SPRING IN THE BATTERY COMPARTMENT GOT DAMAGED BY CORROSION. THE PRODUCT WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
452668 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWB