FDA Adverse Event Other Summary report: N

BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 666784 · Received January 19, 2006

Report

Report Number
2243072-2006-00021
Event Type
Other
Date Received
January 19, 2006
Date of Event
December 19, 2005
Report Date
January 17, 2006
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT HER NEW METER GIVING DIFFERENT READINGS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READINGS (249) AS REFERENCE POINT VS. BD READINGS (324, 285, 207, 191, 177, 311) YIELDED SOME DATA POINTS IN THE D RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064315

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other