FDA Adverse Event
Other
Summary report: N
BD LOGIC BLOOD GLUCOSE MONITOR
MDR report key: 666784
·
Received January 19, 2006
Report
- Report Number
- 2243072-2006-00021
- Event Type
- Other
- Date Received
- January 19, 2006
- Date of Event
- December 19, 2005
- Report Date
- January 17, 2006
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT HER NEW METER GIVING DIFFERENT READINGS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READINGS (249) AS REFERENCE POINT VS. BD READINGS (324, 285, 207, 191, 177, 311) YIELDED SOME DATA POINTS IN THE D RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064315 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |