BONE PIN, 3.2MM X 140MM, STERILE 2 PACK
Report
- Report Number
- 3005985723-2017-00269
- Event Type
- Injury
- Date Received
- June 26, 2017
- Date of Event
- June 8, 2017
- Report Date
- August 31, 2017
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K142530
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: THE REPORTED DEVICE, 3.2MM X 140MM BONE PIN, BROKE DURING INSERTION FOR THE ARRAY PLACEMENT. THE BONE WAS PREDRILLED WITH A 2.0MM DRILL. THE TIP OF THE BONE PIN REMAINS IN THE PATIENT. A SURGICAL DELAY OF 20 MINUTES WAS COMMUNICATED. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION SHOWS THE PIN SHAFT SHEARED OFF APPROXIMATELY 27MM FROM THE START OF THE THREADED SECTION. DIMENSIONAL INSPECTION: THE THREAD MAJOR DIAMETER WAS MEASURED ON THE REMAINING 27MM OF THE PIN SHAFT USING A CALIPER, TN0081-10 (CALIBRATION DUE 31 JULY 2018). THE MEASUREMENT WAS 3.18MM WHICH IS WITHIN THE 3.2±0.05MM TOLERANCE PER DRAWING 143000 REV 00. DEVICE HISTORY REVIEW: A REVIEW OF THE DEVICE HISTORY FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (NON-STERILE P/N 143000-07 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND ACCEPTED ON AUGUST 15TH, 2016. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN STRYKER'S DATABASE RELATED TO P/N 143140, LOT NUMBER W46873 SHOWS NO ADDITIONAL COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE FAILURE WAS CONFIRMED WITH THE EVALUATION OF THE RETURNED DEVICE. A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN TWO CAPAS ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. ONE CAPA HAS BEEN COMPLETED (CATSWEB SYSTEM IS CAPA (B)(4)) WHILE THE SECOND IS STILL OPEN (TRACKWISE (B)(4)). CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
MULTIPLE PINS BROKEN DURING INSERTION FOR THE ARRAY PLACEMENT. SURGEON PREDRILLED WITH A 2.0 DRILL BEFORE PLACING.
MULTIPLE PINS BROKEN DURING INSERTION FOR THE ARRAY PLACEMENT. SURGEON PREDRILLED WITH A 2.0 DRILL BEFORE PLACING
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449342 | BONE PIN, 3.2MM X 140MM, STERILE 2 PACK | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | W46873ROB238 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other |