FDA Adverse Event Injury Summary report: N

BONE PIN, 3.2MM X 140MM, STERILE 2 PACK

MDR report key: 6667195 · Received June 26, 2017

Report

Report Number
3005985723-2017-00269
Event Type
Injury
Date Received
June 26, 2017
Date of Event
June 8, 2017
Report Date
August 31, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K142530
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: THE REPORTED DEVICE, 3.2MM X 140MM BONE PIN, BROKE DURING INSERTION FOR THE ARRAY PLACEMENT. THE BONE WAS PREDRILLED WITH A 2.0MM DRILL. THE TIP OF THE BONE PIN REMAINS IN THE PATIENT. A SURGICAL DELAY OF 20 MINUTES WAS COMMUNICATED. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION SHOWS THE PIN SHAFT SHEARED OFF APPROXIMATELY 27MM FROM THE START OF THE THREADED SECTION. DIMENSIONAL INSPECTION: THE THREAD MAJOR DIAMETER WAS MEASURED ON THE REMAINING 27MM OF THE PIN SHAFT USING A CALIPER, TN0081-10 (CALIBRATION DUE 31 JULY 2018). THE MEASUREMENT WAS 3.18MM WHICH IS WITHIN THE 3.2±0.05MM TOLERANCE PER DRAWING 143000 REV 00. DEVICE HISTORY REVIEW: A REVIEW OF THE DEVICE HISTORY FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (NON-STERILE P/N 143000-07 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND ACCEPTED ON AUGUST 15TH, 2016. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN STRYKER'S DATABASE RELATED TO P/N 143140, LOT NUMBER W46873 SHOWS NO ADDITIONAL COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: THE FAILURE WAS CONFIRMED WITH THE EVALUATION OF THE RETURNED DEVICE. A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN TWO CAPAS ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. ONE CAPA HAS BEEN COMPLETED (CATSWEB SYSTEM IS CAPA (B)(4)) WHILE THE SECOND IS STILL OPEN (TRACKWISE (B)(4)). CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

MULTIPLE PINS BROKEN DURING INSERTION FOR THE ARRAY PLACEMENT. SURGEON PREDRILLED WITH A 2.0 DRILL BEFORE PLACING.

Description of Event or Problem · 1

MULTIPLE PINS BROKEN DURING INSERTION FOR THE ARRAY PLACEMENT. SURGEON PREDRILLED WITH A 2.0 DRILL BEFORE PLACING

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449342 BONE PIN, 3.2MM X 140MM, STERILE 2 PACK STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. W46873ROB238

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other