FDA Adverse Event Malfunction Summary report: N

SYSMEX XN-10

MDR report key: 6665583 · Received June 23, 2017

Report

Report Number
1000515253-2017-00022
Event Type
Malfunction
Date Received
June 23, 2017
Date of Event
February 2, 2017
Report Date
June 23, 2017
Manufacturer
SYSMEX CORPORATION, I SQUARE
Product Code
GKZ
PMA / PMN Number
K112605
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION BY SYSMEX CORPORATION IN (B)(6) DETERMINED LIGHTNING CAUSED AN EXCESSIVE VOLTAGE THAT EXCEEDED THE PRODUCT STANDARD TO BE APPLIED TO THE POWER SUPPLY UNIT. THE SURGE ABSORBER OF THE BOARD WAS DAMAGED. THERE WAS NO ABNORMALITY FOUND IN THE MANUFACTURING OF THE PRODUCT. THE BOARD WAS REPLACED AND THE INSTRUMENT WAS OPERATIONAL. ANY INCIDENT WITH EXCESSIVE HEAT HAS THE POTENTIAL TO CAUSE USER HARM. THE PCB IS MADE OF FLAME RESISTANT MATERIALS AND COVERED WITH A METAL SHIELD, BUT EXCESSIVE HEAT FROM THE EVENT POSES A RISK OF INHALATION OF SMOKE AND/OR HARMFUL VAPORS FROM MELTED MATERIALS. THE PCB IS NOT ACCESSIBLE TO THE USER, AND IS GROUNDED, REDUCING THE POTENTIAL FOR HARM DUE TO SHOCK.

Description of Event or Problem · 1

THE USER REPORTED A POWER OUTAGE OCCURRED DURING A THUNDERSTORM. THE ANALYZER POWER REMAINED ON THROUGH THE USE OF AN UNINTERRUPTED POWER SUPPLY (UPS). WHEN THE USER TRIED TO PERFORM SAMPLE ANALYSIS, THE ANALYZER WAS NOT FUNCTIONING. A SYSMEX FIELD SERVICE ENGINEER (FSE) WENT ON SITE AND FOUND THE SURGE ABSORBER ON THE PRINTED CIRCUIT BOARD (PCB) POWER SUPPLY WAS CHARRED DUE TO EXCESSIVE HEAT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
443254 SYSMEX XN-10 AUTOMATED HEMATOLOGY ANALYZER GKZ SYSMEX CORPORATION, I SQUARE XN-10

Patients

Seq Age Sex Outcome Treatment
1