TRUEMETRIX

MDR report key: 6665319 · Received June 23, 2017

Report

Report Number
1000113657-2017-01080
Event Type
Malfunction
Date Received
June 23, 2017
Date of Event
May 31, 2017
Report Date
June 29, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00010939722447
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT # (B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58 USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP UDI# (B)(4).

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58 USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. DAUGHTER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 198, 435 AND 200 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 240 MG/DL USING TRUEMETRIX METER AND 224 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 04/18/2018 AND OPEN VIAL DATE AT TIME OF CALL IS THREE WEEKS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). MEMORY CONCERNS:CUSTOMERS DAUGHTER IS CONCERNED WITH THE RESULTS OF 198, 435, 200, 241 AND 203 MG/DL IN THE METERS MEMORY.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. DAUGHTER IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 198, 435 AND 200 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 240 MG/DL USING TRUEMETRIX METER AND 224 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 04/18/2018 AND OPEN VIAL DATE AT TIME OF CALL IS THREE WEEKS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
443398 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT2106 00010939722447

Patients

Seq Age Sex Outcome Treatment
1 0 YR