FDA Adverse Event Other Summary report: N

ARCHITECT I2000 ANALYZER

MDR report key: 666481 · Received January 20, 2006

Report

Report Number
1628664-2006-00008
Event Type
Other
Date Received
January 20, 2006
Date of Event
December 23, 2005
Report Date
January 18, 2006
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE ACCOUNT STATES A PT GENERATED ARCHITECT B-HCG RESULTS OF 1.6 AND 1.4 IU/L. THE CONTROLS WERE WITHIN RANGE PRIOR TO RUNNING THE PT SAMPLE. CONTROL RESULTS WERE LATER OUT OF RANGE LOW; THE ACCOUNT REPLACED THE WASH BUFFER AND CONTROLS CAME BACK INTO RANGE. THE PT SAMPLE WAS RETESTED THE FOLLOWING DAY WITH A NEAT RESULT OF >15,000.0 IU/L AND A DILUTED RESULT OF 33,468.2 IU/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000 ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN