ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Report
- Report Number
- 2432235-2017-00372
- Event Type
- Malfunction
- Date Received
- June 23, 2017
- Date of Event
- June 9, 2017
- Report Date
- June 23, 2017
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- GKZ
- PMA / PMN Number
- K102644
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS INC. (SIEMENS) HAS DETERMINED THAT THE CAUSE OF THE NEEDLESTICK INJURY TO THE OPERATOR'S FINGER ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER IS THE OPERATOR'S TECHNIQUE WHEN PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE INSTRUMENT. THE ADVIA 120/ 2120/ 2120I HEMATOLOGY SYSTEM OPERATOR'S GUIDE STATES THE FOLLOWING: "TO AVOID PERSONAL INJURY AND EXPOSURE TO A POTENTIAL BIOHAZARD, YOU MUST COVER THE NEEDLE WITH THE RED NEEDLE COVER IMMEDIATELY AFTER YOU REMOVE THE ENTERING COLLAR." THE OPERATOR FAILED TO FOLLOW THE OPERATOR'S GUIDE INSTRUCTIONS AND DID NOT COVER THE AUTOSAMPLER NEEDLE WITH THE RED NEEDLE STICK COVER IMMEDIATELY AFTER REMOVING THE CENTERING COLLAR. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
AN OPERATOR SUSTAINED A NEEDLESTICK INJURY TO THE FINGER WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE PROCEDURE ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER. THE OPERATOR WENT TO HEALTH SERVICE IN THE SAME HOSPITAL WHERE SHE SUSTAINED HER INJURY. THE OPERATOR REFUSED THE RETROVIRAL TREATMENT OFFERED BY THE HOSPITAL. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE OPERATOR SUSTAINING A NEEDLESTICK INJURY WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE PROCEDURE ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 443935 | ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER | ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER | GKZ | SIEMENS HEALTHCARE DIAGNOSTICS INC. | ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |