FDA Adverse Event Malfunction Summary report: N

ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER

MDR report key: 6662785 · Received June 23, 2017

Report

Report Number
2432235-2017-00372
Event Type
Malfunction
Date Received
June 23, 2017
Date of Event
June 9, 2017
Report Date
June 23, 2017
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
GKZ
PMA / PMN Number
K102644
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS INC. (SIEMENS) HAS DETERMINED THAT THE CAUSE OF THE NEEDLESTICK INJURY TO THE OPERATOR'S FINGER ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER IS THE OPERATOR'S TECHNIQUE WHEN PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE INSTRUMENT. THE ADVIA 120/ 2120/ 2120I HEMATOLOGY SYSTEM OPERATOR'S GUIDE STATES THE FOLLOWING: "TO AVOID PERSONAL INJURY AND EXPOSURE TO A POTENTIAL BIOHAZARD, YOU MUST COVER THE NEEDLE WITH THE RED NEEDLE COVER IMMEDIATELY AFTER YOU REMOVE THE ENTERING COLLAR." THE OPERATOR FAILED TO FOLLOW THE OPERATOR'S GUIDE INSTRUCTIONS AND DID NOT COVER THE AUTOSAMPLER NEEDLE WITH THE RED NEEDLE STICK COVER IMMEDIATELY AFTER REMOVING THE CENTERING COLLAR. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 1

AN OPERATOR SUSTAINED A NEEDLESTICK INJURY TO THE FINGER WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE PROCEDURE ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER. THE OPERATOR WENT TO HEALTH SERVICE IN THE SAME HOSPITAL WHERE SHE SUSTAINED HER INJURY. THE OPERATOR REFUSED THE RETROVIRAL TREATMENT OFFERED BY THE HOSPITAL. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE OPERATOR SUSTAINING A NEEDLESTICK INJURY WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE PROCEDURE ON THE ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
443935 ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER GKZ SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER

Patients

Seq Age Sex Outcome Treatment
1