FDA Adverse Event Malfunction Summary report: N

OAKWORKS MEDICAL CFPM 400 IMAGING TABLE

MDR report key: 6660011 · Received June 22, 2017

Report

Report Number
2529571-2017-00006
Event Type
Malfunction
Date Received
June 22, 2017
Date of Event
March 18, 2015
Report Date
June 6, 2017
Manufacturer
OAKWORKS, INC.,
Product Code
IXR
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RECURRING ISSUE.

Description of Event or Problem · 1

THE END USER CLINIC IS HAVING PROBLEMS WITH THE FOOT CONTROL. WHEN THEY WALK AROUND THE TABLE, THEY ARE KICKING THE FOOT CONTROL UP UNDER THE LOWER FRAME AND THE UP FUNCTION IS ENGAGING WITH THE PATIENT ON THE TABLE. THE NURSES ARE AWARE THAT THIS CAN AND DOES HAPPEN. THEY WERE ADVISED TO UNPLUG THE FOOT PEDAL AND USE ONLY THE HAND CONTROL. THE CUSTOMER STATED THAT NO PATIENT HARM HAD OCCURRED AS THEY WERE AWARE OF WHAT CAUSED THE MOVEMENT OF THE TABLE AND WERE ABLE TO REACT QUICKLY.