FDA Adverse Event
Malfunction
Summary report: N
OAKWORKS MEDICAL CFPM 400 IMAGING TABLE
MDR report key: 6659993
·
Received June 22, 2017
Report
- Report Number
- 2529571-2017-00003
- Event Type
- Malfunction
- Date Received
- June 22, 2017
- Date of Event
- August 29, 2014
- Report Date
- June 5, 2017
- Manufacturer
- OAKWORKS, INC.,
- Product Code
- IXR
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
THE FOOT CONTROL BUTTON WAS STUCK AND CAUSED THE TABLE TO MOVE INTO A LATERAL ROLL WHEN THEY WERE TRYING TO LOWER THE TABLE. THE PATIENT WAS ON THE TABLE AT THE TIME AS THE PROCEDURE HAD CONCLUDED AND THEY WERE ATTEMPTING TO GET THE PATIENT OFF THE TABLE WHEN THE UNINTENDED MOVEMENT OCCURRED. THE TOP WENT TO A 45 DEGREE ANGLE BEFORE THE UNINTENDED MOVEMENT STOPPED. IT TOOK THREE PEOPLE TO GET THE PATIENT OFF THE TABLE SAFELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441891 | OAKWORKS MEDICAL CFPM 400 IMAGING TABLE | IMAGING TABLE | IXR | OAKWORKS, INC., | CFPM 400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |