FDA Adverse Event
Malfunction
Summary report: N
OAKWORKS MEDICAL CFPM 401 IMAGING TABLE
MDR report key: 6659962
·
Received June 22, 2017
Report
- Report Number
- 2529571-2017-00001
- Event Type
- Malfunction
- Date Received
- June 22, 2017
- Date of Event
- August 14, 2014
- Report Date
- June 5, 2017
- Manufacturer
- OAKWORKS, INC.,
- Product Code
- IXR
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NOT RETURNED.
Description of Event or Problem · 1
THE COMPLAINT WAS RECEIVED BY OAKWORKS, INC., BY A CUSTOMER IN (B)(6). THE PHYSICIAN HAD VISITED AN OAKWORKS, INC., BOOTH AT A TRADE SHOW. HE EXPLAINED THAT HE HAD TRIALED A CFPM 401 IMAGING TABLE AND WANTED TO PROVIDE FEEDBACK. THERE WAS NO PATIENT INVOLVEMENT AND NO PROCEDURE. HE MERELY SHOWED THE OAKWORKS, INC. REPRESENTATIVE ON THE TABLE IN THE BOOTH AT THE TRADE SHOW THAT THE FOOD PEDAL CAN SLIP UNDER THE TABLE, CAUSING ONE OF THE POWER MOTIONS TO BE ACTIVATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442690 | OAKWORKS MEDICAL CFPM 401 IMAGING TABLE | IMAGING TABLE | IXR | OAKWORKS, INC., | CFPM 401 | NOT LISTED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |