I-STAT DOWNLOADER, NETWORK

MDR report key: 6646195 · Received June 16, 2017

Report

Report Number
2245578-2017-00025
Event Type
Malfunction
Date Received
June 16, 2017
Date of Event
June 6, 2017
Report Date
September 15, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
RHA
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09/08/2017. THE FAILURE WAS DUE TO A FAILED TRANSISTOR Q1 ON THE FAST CHARGE PCBA OF THE DRN.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

NA.

Description of Event or Problem · 1

ON (B)(6) 2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED VISUAL SMOKE COMING FROM NETWORK DOWNLOADER (DRN-19611). THE DEVICE WAS TAKEN OUT OF SERVICE, REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
429837 I-STAT DOWNLOADER, NETWORK DOWNLOADER, NETWORK RHA ABBOTT POINT OF CARE NA NA

Patients

Seq Age Sex Outcome Treatment