FLEXLAB
Report
- Report Number
- 3010825766-2017-00003
- Event Type
- Malfunction
- Date Received
- June 16, 2017
- Date of Event
- May 17, 2017
- Report Date
- June 16, 2017
- Manufacturer
- INPECO SA
- Product Code
- JQP
- PMA / PMN Number
- K121012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
THE EVENT WAS INVESTIGATED BY INPECO, WHICH DETERMINED THAT THE ABNORMAL NUMBER OF CLOT DETECTION ERRORS AND THE SAMPLES DILUTION WERE DUE TO AN INCORRECT MAINTENANCE OF THE ALIQUOTER UNIT. SPECIFICALLY, THE USER DID NOT USE DISTILLED OR DE-IONIZED WATER AS CLEARLY INDICATED IN THE SERVICE MANUAL. AS A CONSEQUENCE, ALIQUOTER UNIT COMPONENTS WERE DAMAGED. AS CORRECTIVE ACTION THE DISTRIBUTOR HAS REPLACED THE DAMAGED ALIQUOTER UNIT COMPONENTS. THE RELEVANT PART OF THE SERVICE MANUAL HAS BEEN REVIEWED AND IT HAS BEEN CONFIRMED THE INSTRUCTIONS ARE CLEAR AND NOT MISLEADING. THE PRODUCT DESIGN AND LABELING ARE CONSIDERED ADEQUATE. NO PRODUCT CHANGE FORESEEN. THIS DEVICE IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
INPECO WAS NOTIFIED BY THE FLEXLAB DISTRIBUTOR ABOUT AN ISSUE WITH THE ALIQUOTER UNIT. SPECIFICALLY, A SMALL VOLUME OF WATER FROM THE ALIQUOTER PIPETTOR CIRCUIT WAS DISPENSED INTO SAMPLE TUBES DILUTING THEM. THE DISTRIBUTOR REFERRED ALSO THAT THE CUSTOMER EXPERIENCED AN UNEXPECTED INCREASED NUMBER OF ERRORS (I.E. CLOT DETECTION ERROR) THE DAYS BEFORE THE EVENT. AS IMMEDIATE ACTION THE DISTRIBUTOR SET THE ALIQUOTER MODULE OFF-LINE. INPECO REVIEWED THE LOG FILES AND IDENTIFIED THE SAMPLE TUBES PROCESSED BY ALIQUOTER POTENTIALLY AFFECTED BY DILUTION. THE LIST OF SAMPLES ID WAS COMMUNICATED TO THE CUSTOMER IN ORDER TO LET THEM DOUBLE CHECK THE TESTS RESULT. THERE WAS NO DEATH NEITHER SERIOUS INJURY TO ANY PATIENT DUE TO ERRONEOUS TEST RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 426792 | FLEXLAB | LABORATORY AUTOMATION SYSTEM | JQP | INPECO SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |